SALVE Cc Correction Realized
August 9, 2016 – With a new corrector for both spherical and chromatic aberration, the SALVE III microscope resolves atoms at every voltage from 20 to 80 kV. Scientists from CEOS, FEI (today Thermo Fisher Scientific) and Ulm University reached an information limit of 90 pm at 40 kV, only 15 times the electron wavelength and a record for phase-contrast TEM, and imaged single atoms in graphene and MoS2 at 30 kV.
Seeing single atoms in a transmission electron microscope (TEM) is routine today, as long as the sample can take the beam. Many of the most interesting materials cannot: graphene, molybdenum disulfide and other light-atom sheets lose atoms when struck by fast electrons. Slowing the electrons down protects the sample but, until recently, blurred the image. Scientists from CEOS in Heidelberg, FEI Company in Eindhoven (today part of Thermo Fisher Scientific) and Ulm University have now put into operation a new corrector that removes both the spherical (Cs) and the chromatic (Cc) aberration of the objective lens at electron energies from 20 to 80 keV. With it, the SALVE III microscope resolves atoms at all these voltages. Its best resolution of 90 pm at 40 kV is only 15 times the electron wavelength, a record for phase-contrast TEM imaging (Fig. 1).
Why low voltages need a new corrector
Since the beginning of aberration-corrector technology, TEM has provided sub-Ångström instrumental resolution, provided that the specimen is not damaged by the electron beam[1],[2]. To avoid knocking atoms out of their places, the energy of the electrons must stay below a threshold that grows with the mass of the atom[3]. Specimens made of light elements, such as carbon materials, oxides and lithium compounds, suffer such knock-on damage already at the usual energies[4]–[6], and biological objects, polymers and ceramics are damaged by ionization as well, even below the knock-on threshold.
To avoid knock-on damage at atomic resolution for many materials, an aberration-corrected microscope has to work at 80 kV or less. The first such TEM was completed in 2005[7], and with it exposure times could be increased considerably, opening atomic resolution to a whole range of objects. Experiments soon showed, however, that interfaces and defects in light-element materials, which are what researchers most want to see, need even lower voltages, because their atoms are displaced more easily. To reach atomic resolution for these materials, the SALVE project (Sub-Ångström Low-Voltage Electron microscopy)[8] was launched in 2009. Its microscope was to work between 20 and 80 kV, below the knock-on threshold of the object.
At such voltages, correcting the spherical aberration alone is not enough. Without Cc correction, the resolution stays at about 0.2 nm[9], because electrons of slightly different energy are focused at different heights, and the slower the electrons, the larger this effect. A comparison at 20 kV makes the difference plain (Fig. 1): in an uncorrected TEM and in a Cs-only corrected TEM, small gold clusters on a thin carbon film appear as dark blobs, whereas only the Cc/Cs-corrected SALVE microscope resolves them atom by atom, with the gold lattice spacing of 144 pm clearly visible in the diffractogram.
The Rose–Kuhn corrector
“We were able to achieve the new resolution record due to our latest understanding of resolution-limiting factors in electron microscopy and by applying the improved understanding for the optimization of the corrector,” said Prof. Max Haider, director of CEOS GmbH, the company that develops and builds aberration correctors.
The SALVE corrector is based on a design that Prof. Harald Rose and his student Mathias Kuhn proposed in 1992[10]. Rose, who advised the SALVE project, had originally suggested it for an electron-optical aplanat, a system free of the primary chromatic, spherical and off-axial aberrations, exactly as the SALVE microscope required. During the evaluation phase (SALVE I, 2009–2011), the researchers compared all corrector designs under consideration and concluded that this one offered by far the best performance. “I had completely forgotten this design over the years because the simultaneous correction of these aberrations was not necessary at the standard voltages above 80 kV used in TEMs so far,” said Rose.
In SALVE II (2011–2014), the corrector was developed, and in SALVE III (2015–2016) it was optimized and integrated into the microscope. The corrector adds 43 cm of length and about 185 kg of weight to the column and consists of eight multipole elements (Fig. 2). “That the third phase could be so successfully completed is also due to the enormous commitment of SALVE III partner FEI Company which supplied and developed the microscope column,” said Prof. Ute Kaiser, principal investigator of the SALVE project. How the corrector works is described in a box below.
What limits resolution
In the years before, the team at CEOS had greatly improved the understanding of what limits the resolution of a TEM. Three parameters set the information limit: the spatial coherence of the illumination, the focus spread in the imaging optics caused by unstable electromagnetic fields, and an image spread caused by Johnson noise[13]. Correcting the spherical aberration removes most of the damping effect of partial spatial coherence; that Cs correction improves the resolution of an electron microscope was first demonstrated in 1998[1].
The chromatic aberration turns the energy spread of the electron source and the instabilities of the accelerating voltage and lens fields into a focus spread. This focus spread is the main limitation for low-voltage TEM imaging, whether the spherical aberration is corrected or not[14]. Besides the primary chromatic aberration, the higher-order chromatic aberrations must be kept small at low voltages as well, as had been shown in 2010[15]. Even after Cc correction, a focus spread remains from noise in the electronic supplies of the focusing elements. It can only be reduced with ultra-stable voltage and current sources[16] and a room that shields the microscope from mechanical and electromagnetic disturbances.
Finally, every conductor near the beam, magnetic or not, generates tiny random magnetic fields, the so-called Johnson noise, which deflect the electrons and smear out the image[13]. To keep this image spread small, the ray path and the shape of the metal parts close to the beam were optimized. The bore of the multipoles, the lens gaps and the diameter of the liner tube had to be balanced so that the corrector provided the required correction strength and focusing power without exposing the beam to too much noise. The design went to the edge of what is technically feasible: the magnetic materials in the multipoles, for example, were driven close to saturation.
A record resolution
The measurements showed that the corrector works over the whole voltage range. The researchers report an information limit between 76 pm at 80 kV and 139 pm at 20 kV (Table 1)[17]. They determined it with Young’s fringes, interference patterns of two slightly shifted images of a 2 nm thin amorphous tungsten film (Fig. 3), and confirmed it with an independent test, the so-called paraboloid method, which evaluates focal series in three-dimensional Fourier space[17]–[20]. Both tests agreed and showed that the SALVE goal, a usable aperture of 50 mrad, was reached and even exceeded at every voltage.
| Voltage | Wavelength (pm) | Lattice spacing at 50 mrad (pm) | Information limit (pm) | Equivalent aperture (mrad) |
|---|---|---|---|---|
| 20 kV | 171.8 | 8.589 | 139 | 62 |
| 30 kV | 139.6 | 6.979 | 115 | 61 |
| 40 kV | 120.3 | 6.016 | 90 | 67 |
| 60 kV | 97.3 | 4.866 | 83 | 59 |
| 80 kV | 83.5 | 4.176 | 76 | 55 |
Table 1: Measured information limit of the SALVE microscope and its equivalent aperture at five voltages, compared with the lattice spacing that a 50 mrad aperture would resolve (Table I of the paper).
To make full use of this information, the aberrations must be tuned precisely across the whole aperture. The round aberrations, defocus C1 and the spherical aberrations of third order C3 and fifth order C5, were set to specific values to give a uniform phase contrast transfer function, and all non-round aberrations were reduced so far that they no longer disturbed the phase contrast[21]. The measured phase plate at 20 kV shows that this demanding task was accomplished (Fig. 3f). The off-axial aberrations, including the off-axial coma, were also tuned small enough to image every point of a large field of view in the same way. The results will also help to improve Cc correctors designed before the resolution-limiting effect of Johnson noise was discovered. Why chromatic correction matters so much at low voltage, and how stable the SALVE microscope turned out to be, is explained in a box below (Fig. 4).
Single atoms at 30 kV
With the new corrector, the resolution for graphene and MoS2 at 30 kV is comparable to the state of the art at 80 kV. This is owed both to the large usable aperture of the instrument and to the stronger scattering of slower electrons. Images of free-standing monolayers of graphene and MoS2 at 30 kV show a strongly improved single-atom contrast that matches image simulations (Fig. 5). In the raw images, taken with 1 s exposure and 2048 × 2048 pixels, the accumulated electron doses were 2.5 × 106 e−/nm2 for graphene and 3 × 105 e−/nm2 for MoS2.
The diffractogram of graphene showed reflections up to third order, corresponding to 107 pm; for MoS2, even the fourth-order reflections at 102 pm were detected. Although the raw images are very noisy at these limited doses, the single carbon atoms of the hexagonal lattice can be identified, and molybdenum can clearly be told apart from the two sulfur atoms that sit on top of each other in this projection. Averaging raised the signal-to-noise ratio and gave images in good agreement with the simulations[17]. If the agreement between experiment and calculation is developed further into a truly quantitative match, the authors write, it “will not only allow for detecting single atoms and point-defect structures but also for determining their chemical nature.”
The gentler beam also pays off in time-resolved imaging. With the SALVE III microscope, an edge configuration in graphene could be imaged at atomic resolution for the first time: it stays fixed at 30 kV, whereas it is reproducibly unstable under an 80 kV beam[25],[26] (Video 1).
Video 1: An edge configuration in graphene imaged with the SALVE III microscope at 30 kV. It hardly changes during the recording, whereas the same configuration is reproducibly unstable at 80 kV (Supplemental Movie of the paper).
Another route: a monochromator instead of Cc correction
In the same years, JEOL and AIST in Japan explored the other way to fight the chromatic blur at low voltage: correcting the geometrical aberrations up to fifth order and narrowing the energy spread of the electrons, without correcting Cc. Their last study was submitted in June 2016, two months before the SALVE results appeared, so both groups show the state of low-voltage TEM at the same time. The first box below holds the 15 kV tests of the JEOL Triple C microscope, with its Young’s fringes and h-BN images (Figs. 6 and 7).
JEOL then added a monochromator and reported atomic resolution in graphene at 60 and 15 kV; the second box checks the resolution claimed for these images.
Fig. 8 puts side by side the information limits measured with Young’s fringes. At 20 kV, the Cc/Cs corrector improved the resolution of the SALVE microscope from 200 pm to 139 pm[14], while the JEOL Triple C #1 without Cc correction reached 200 pm at 15 kV. Of the monochromated JEOL TEM, only what its images prove is shown: the bright 213 pm reflections of graphene at 15 kV, which lie well inside the transferred range and are therefore a reliable upper bound for the resolution. It is no better than the 200 pm that Triple C #1 reached without a monochromator. The finer values claimed for this microscope are left out, because they rest on fitted diffractograms and on non-linear spots, not on a direct measurement of the linear transfer.
A big step forward for all sciences that use sensitive materials
The scientists from CEOS, FEI and Ulm University demonstrated a TEM corrector for spherical and chromatic aberration that works from 20 to 80 kV. Between 40 and 80 kV it reaches sub-Ångström resolution, and the information limit of 90 pm at 40 kV corresponds to a maximum aperture angle of 67 mrad. At 30 kV, microscope and corrector together resolve single atoms in raw images, and even at 20 kV, the lowest voltage available, the resolution is better than 0.14 nm. Because energy windows of many electronvolts are focused almost equally well, chromatic correction also opens new possibilities for analytical imaging. “This is a huge step forward towards materials analysis on delicate, beam-sensitive objects,” the authors conclude; “the SALVE microscope now combines both the desired resolution and the beam energy below knockon threshold at the same time.”
Choosing the right voltage remains a balance. As the voltage drops, ionization damage, known as radiolysis, increases, so the SALVE project also worked on reducing it through better sample preparation[29],[30]. The best electron energy depends on each sample’s compromise between ionization and knock-on damage[31], and finding the optimum voltage for different materials and scientific questions was another aim of the project. Today the SALVE microscope is part of the SALVE Microscopy Center at Ulm University.
“With the new SALVE III microscope scientists can now examine for the first time the sample with atomic resolution at electron energies in the range between 20 and 80 keV. This possibility is a great step for electron microscopists and all sciences which are using electron microscopy,” said Ute Kaiser. More about the contributions of Otto Scherzer and Harald Rose to hardware aberration correction can be found in the SALVE news Scherzer’s and Rose’s contributions.
Resource:
Linck, M., Hartel, P., Uhlemann, S., Kahl, F., Müller, H., Zach, J., Haider, M., Niestadt, M., Bischoff, M., Biskupek, J., Lee, Z., Lehnert, T., Börrnert, F., Rose, H., Kaiser, U. (2016).
Chromatic Aberration Correction for Atomic Resolution TEM Imaging from 20 to 80 kV.
Phys. Rev. Lett. 117(7), 076101.
https://doi.org/10.1103/PhysRevLett.117.076101
Update – August 30, 2018
The news article of August 9, 2016 ends above. Two years later, the SALVE team tested the corrected microscope on graphene at both ends of its voltage range.
Graphene at 80 kV and 20 kV
Felix Börrnert and Ute Kaiser imaged single-layer graphene with the SALVE III microscope at 80 and 20 kV and calculated, for each image, the phase contrast transfer function from the aberrations measured on the instrument[32]. At 80 kV, the raw image, taken with 2.3 × 107 e−/nm2 in 2 s, separated every carbon atom with a contrast of about 10 %, and its diffractogram showed signals out to 71 pm. The calculation gave a point resolution of 78 pm and an information limit of 75 pm, in line with the 76 pm measured with Young’s fringes in 2016 (Table 4). The strongest damping came from image spread caused by Johnson noise[13], at most 22.5 pm. Because image spread also damps non-linear transfer, nearly all the information in a SALVE image lies within the information limit. Without Cc correction this is different: an 80 kV Titan without monochromator has an information limit of 190 pm[19], but was often credited with 120 pm because of non-linear second-order graphene spots in its diffractograms.
In a Fourier-filtered version of the 80 kV image, in which the perfect lattice was removed, one point defect stood out: a single vacancy whose surroundings were clearly asymmetric. This Jahn–Teller distortion, caused by the electronic structure of the vacancy, had to the authors’ knowledge never been imaged directly at true atomic resolution before.
At 20 kV, each pixel of the raw image held a signal of only about 40 electrons on a background of 700. Since the image was oversampled more than eight times, binning four pixels by four raised the signal 16-fold without losing resolution, and the carbon atoms, 140 pm apart, were clearly resolved with about 7 % contrast. The information limit was at least 130 pm, a 64 mrad wide open phase plate, or about 15.5 times the electron wavelength. The point resolution of 150 pm, however, was set by the remaining fifth-order spherical aberration C5 of 6.7 mm. It also caused a delocalization of about 180 pm at the graphene lattice frequency, more than the atom spacing, so a “phantom” atom could appear in a vacancy. This explained a surprise: a single vacancy showed up only in black-atom contrast, at a defocus about 5 nm away, where the delocalization for graphene drops to zero. A heavier atom at the edge of a hole stayed bright even when the graphene lattice reversed its contrast, a sign that the weak phase object approximation fails for heavier atoms at low voltage. Under the beam, the hole grew by roughly one unit cell per edge per 108 e−/nm2.
The same C5 makes the depth of focus small. At 20 kV, the whole pass band stayed positive only within about 1 nm of defocus. A simple test confirmed the scale: graphene tilted by ±30° showed bands of strongest lattice contrast whose spacing corresponds to 5.2 nm of defocus, against 5.3 nm calculated. With C5 reduced to 0.5 mm, the calculation shows straight transfer bands and a delocalization below 100 pm.
The authors also compare the other routes to low-voltage resolution. A monochromator removes most of the electrons, and the inelastically scattered ones then only add noise; with Cc correction the whole beam is used, and inelastic electrons still contribute to the image. Extremely narrow pole-piece gaps[24] halve the aberrations but leave little room for the sample, need flat holders and limit tilting. SALVE III keeps a standard 5.4 mm gap that takes every commercial in situ holder. “Chromatic-aberration correction in addition to geometric-aberration correction in low-voltage TEM is greatly beneficial for in situ imaging with atomic resolution,” the authors conclude.
| SALVE III 2016: Young’s fringes [17] | SALVE III 2018: graphene, calculated transfer [32] | Agreement | |
|---|---|---|---|
| Information limit, 80 kV | 76 pm | 75 pm; diffractogram signals to 71 pm | confirmed |
| Point resolution, 80 kV | – | 78 pm | new |
| Information limit, 20 kV | 139 pm | at least 130 pm (64 mrad) | confirmed |
| Point resolution, 20 kV | – | 150 pm, set by C5 = 6.7 mm | new limit |
| Focus spread | ≤ 1000 pm at 20 kV | ≤ 500 pm at 80 kV | confirmed |
| Image spread (Johnson noise) | dominant limit [13] | 22.5 pm at 80 kV, 39 pm at 20 kV, dominant | confirmed |
| Graphene | single atoms in raw images at 30 kV | atoms 140 pm apart resolved in a raw image at 20 kV | extended to 20 kV |
| Delocalization, 20 kV | – | about 180 pm; single vacancy visible only in black-atom contrast | new: smaller C5 needed |
Table 4: SALVE III in 2016 (this article) and in the 2018 graphene study (ref. [32]).
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Haider, M., Uhlemann, S., Schwan, E., Rose, H., Kabius, B., Urban, K. (1998). Electron Microscopy Image Enhanced. Nature, 392, 768–769. https://doi.org/10.1038/33823
-
Kisielowski, C., Freitag, B., Bischoff, M., van Lin, H., Lazar, S., Knippels, G., Tiemeijer, P., van der Stam, M., von Harrach, S., Stekelenburg, M., Haider, M., Uhlemann, S., Müller, H., Hartel, P., Kabius, B., Miller, D., Petrov, I., Olson, E. A., Donchev, T., Kenik, E. A., Lupini, A. R., Bentley, J., Pennycook, S. J., Anderson, I. M., Minor, A. M., Schmid, A. K., Duden, T., Radmilovic, V., Ramasse, Q. M., Watanabe, M., Erni, R., Stach, E. A., Denes, P., Dahmen, U. (2008). Detection of Single Atoms and Buried Defects in Three Dimensions by Aberration-Corrected Electron Microscope With 0.5-Å Information Limit. Microsc. Microanal., 14, 469–477. https://doi.org/10.1017/S1431927608080902
-
Scherzer, O. (1970). Die Strahlenschädigung der Objekte als Grenze für die hochauflösende Elektronenmikroskopie. Ber. Bunsenges. Phys. Chem., 74, 1154–1167. https://doi.org/10.1002/bbpc.19700741113
-
Banhart, F. (1999). Irradiation Effects in Carbon Nanostructures. Rep. Prog. Phys., 62, 1181–1221. https://doi.org/10.1088/0034-4885/62/8/201
-
Meyer, J. C., Eder, F., Kurasch, S., Skakalova, V., Kotakoski, J., Park, H. J., Roth, S., Chuvilin, A., Eyhusen, S., Benner, G., Krasheninnikov, A. V., Kaiser, U. (2012). Accurate Measurement of Electron Beam Induced Displacement Cross Sections for Single-Layer Graphene. Phys. Rev. Lett., 108, 196102. https://doi.org/10.1103/PhysRevLett.108.196102
-
Wang, F., Graetz, J., Moreno, M. S., Ma, C., Wu, L., Volkov, V., Zhu, Y. (2011). Chemical Distribution and Bonding of Lithium in Intercalated Graphite: Identification With Optimized Electron Energy Loss Spectroscopy. ACS Nano, 5, 1190–1197. https://doi.org/10.1021/nn1028168
-
Kujawa, S., Freitag, B., Hubert, D. (2005). An Aberration Corrected (S)TEM Microscope for Nanoresearch. Microscopy Today, 13, 16–21. https://doi.org/10.1017/S1551929500053608
-
Kaiser, U. (2016). Project homepage of the Sub-Ångström Low-Voltage Electron Microscopy and Spectroscopy (SALVE) project. Ulm University, Germany, www.salve-project.de.
-
Alem, N., Yazyev, O. V., Kisielowski, C., Denes, P., Dahmen, U., Hartel, P., Haider, M., Bischoff, M., Jiang, B., Louie, S. G., Zettl, A. (2011). Probing the Out-of-Plane Distortion of Single Point Defects in Atomically Thin Hexagonal Boron Nitride at the Picometer Scale. Phys. Rev. Lett., 106, 126102. https://doi.org/10.1103/PhysRevLett.106.126102
-
Rose, H. (1992). Correction of Aberrations – A Promising Method for Improving the Performance of Electron Microscopes. In: Electron Microscopy 92, Proceedings of the 10th European Congress on Electron Microscopy, Granada, Spain, 7–11 September 1992. Secretariado de Publicaciones de la Universidad de Granada, Granada.
-
Scherzer, O. (1947). Sphärische und chromatische Korrektur von Elektronen-Linsen. Optik, 2, 114–132.
-
Jia, C. L., Lentzen, M., Urban, K. (2004). High-Resolution Transmission Electron Microscopy Using Negative Spherical Aberration. Microsc. Microanal., 10, 174–184. https://doi.org/10.1017/S1431927604040425
-
Uhlemann, S., Müller, H., Hartel, P., Zach, J., Haider, M. (2013). Thermal Magnetic Field Noise Limits Resolution in Transmission Electron Microscopy. Phys. Rev. Lett., 111, 046101. https://doi.org/10.1103/PhysRevLett.111.046101
-
Kaiser, U., Biskupek, J., Meyer, J. C., Leschner, J., Lechner, L., Rose, H., Stöger-Pollach, M., Khlobystov, A. N., Hartel, P., Müller, H., Haider, M., Eyhusen, S., Benner, G. (2011). Transmission Electron Microscopy at 20 kV for Imaging and Spectroscopy. Ultramicroscopy, 111, 1239–1246. https://doi.org/10.1016/j.ultramic.2011.03.012
-
Haider, M., Hartel, P., Müller, H., Uhlemann, S., Zach, J. (2010). Information Transfer in a TEM Corrected for Spherical and Chromatic Aberration. Microsc. Microanal., 16, 393–408. https://doi.org/10.1017/S1431927610013498
-
Haider, M., Müller, H., Uhlemann, S., Zach, J., Loebau, U., Hoeschen, R. (2008). Prerequisites for a Cc/Cs-Corrected Ultrahigh-Resolution TEM. Ultramicroscopy, 108, 167–178. https://doi.org/10.1016/j.ultramic.2007.07.007
-
Linck, M. et al. (2016). Supplemental Material to “Chromatic Aberration Correction for Atomic Resolution TEM Imaging From 20 to 80 kV”: Aberration Coefficients, an Alternative Resolution Assessment, Image Processing Details and a Description of the Supplemental Movie. Phys. Rev. Lett., 117, 076101, link.aps.org/supplemental/10.1103/PhysRevLett.117.076101.
-
Müller, H., Maßmann, I., Uhlemann, S., Hartel, P., Zach, J., Haider, M. (2011). Aplanatic Imaging Systems for the Transmission Electron Microscope. Nucl. Instrum. Methods Phys. Res. A, 645, 20–27. https://doi.org/10.1016/j.nima.2010.12.091
-
Barthel, J., Thust, A. (2008). Quantification of the Information Limit of Transmission Electron Microscopes. Phys. Rev. Lett., 101, 200801. https://doi.org/10.1103/PhysRevLett.101.200801
-
Kimoto, K., Kurashima, K., Nagai, T., Ohwada, M., Ishizuka, K. (2012). Assessment of Lower-Voltage TEM Performance Using 3D Fourier Transform of Through-Focus Series. Ultramicroscopy, 121, 31–37. https://doi.org/10.1016/j.ultramic.2012.06.012
-
Hartel, P., Linck, M., Kahl, F., Müller, H., Haider, M. (2014). On Proper Phase Contrast Imaging in Aberration Corrected TEM. Microsc. Microanal., 20, 926–927. https://doi.org/10.1017/S1431927614006357
-
Kirkland, A. I., Saxton, W. O., Chau, K. L., Tsuno, K., Kawasaki, M. (1995). Super-Resolution by Aperture Synthesis: Tilt Series Reconstruction in CTEM. Ultramicroscopy, 57, 355–374. https://doi.org/10.1016/0304-3991(94)00191-O
-
Sasaki, T., Sawada, H., Hosokawa, F., Sato, Y., Suenaga, K. (2014). Aberration-Corrected STEM/TEM Imaging at 15 kV. Ultramicroscopy, 145, 50–55. https://doi.org/10.1016/j.ultramic.2014.04.006
-
Morishita, S., Mukai, M., Suenaga, K., Sawada, H. (2016). Resolution Enhancement in Transmission Electron Microscopy With 60-kV Monochromated Electron Source. Appl. Phys. Lett., 108, 013107. https://doi.org/10.1063/1.4939653
-
Kotakoski, J., Santos-Cottin, D., Krasheninnikov, A. V. (2012). Stability of Graphene Edges Under Electron Beam: Equilibrium Energetics versus Dynamic Effects. ACS Nano, 6, 671–676. https://doi.org/10.1021/nn204148h
-
Suenaga, K., Iizumi, Y., Okazaki, T. (2011). Single Atom Spectroscopy With Reduced Delocalization Effect Using a 30 kV-STEM. Eur. Phys. J. Appl. Phys., 54, 33508. https://doi.org/10.1051/epjap/2011100414
-
Morishita, S., Mukai, M., Suenaga, K., Sawada, H. (2016). Atomic Resolution Imaging at an Ultralow Accelerating Voltage by a Monochromatic Transmission Electron Microscope. Phys. Rev. Lett., 117, 153004 (received 20 June 2016). https://doi.org/10.1103/PhysRevLett.117.153004
-
Bell, D. C., Russo, C. J., Kolmykov, D. V. (2012). 40 keV Atomic Resolution TEM. Ultramicroscopy, 114, 31–37. https://doi.org/10.1016/j.ultramic.2011.12.001
-
Algara-Siller, G., Kurasch, S., Sedighi, M., Lehtinen, O., Kaiser, U. (2013). The Pristine Atomic Structure of MoS2 Monolayer Protected From Electron Radiation Damage by Graphene. Appl. Phys. Lett., 103, 203107. https://doi.org/10.1063/1.4830036
-
Chamberlain, T. W., Biskupek, J., Skowron, S. T., Bayliss, P. A., Bichoutskaia, E., Kaiser, U., Khlobystov, A. N. (2015). Isotope Substitution Extends the Lifetime of Organic Molecules in Transmission Electron Microscopy. Small, 11, 622–629. https://doi.org/10.1002/smll.201402081
-
Egerton, R. F. (2014). Choice of Operating Voltage for a Transmission Electron Microscope. Ultramicroscopy, 145, 85–93. https://doi.org/10.1016/j.ultramic.2013.10.019
-
Börrnert, F., Kaiser, U. (2018). Chromatic- and Geometric-Aberration-Corrected TEM Imaging at 80 kV and 20 kV. Phys. Rev. A, 98, 023861. https://doi.org/10.1103/PhysRevA.98.023861









