Thermal Magnetic Noise
July 22, 2013 – Researchers at CEOS in Heidelberg, the corrector partner of the SALVE project, have found what limited the resolution of the best aberration-corrected electron microscopes: magnetic field noise from thermally driven currents in the metal around the beam. In a Titan at 80 kV, cooling a test tube in the beam path with liquid nitrogen clearly sharpened the images.
Aberration correctors had pushed the transmission electron microscope (TEM) to ever finer detail, yet the newest instruments still fell short of what their optics promised, and nobody knew why; as Nature News put it, imaging had hit a noise barrier[1]. With correctors for both the spherical and the chromatic aberration, linear contrast transfer had reached spacings of 80 pm at 80 kV[2], but the information limit, the finest detail an instrument transfers, stayed clearly behind the value its residual aberrations and instabilities allowed. Stephan Uhlemann, Heiko Müller, Peter Hartel, Joachim Zach and Max Haider at CEOS in Heidelberg, the corrector partner of the SALVE project, have now found the cause: tiny magnetic fields from thermally driven currents in every piece of metal around the beam, the magnetic counterpart of the Johnson noise known from electronics. In a modified FEI Titan at 80 kV they replaced the corrector by a test tube that could be cooled with liquid nitrogen, and the blur in the images shrank as the tube got colder. Since only one microscope was involved, a box below sets the measurements against the theory of thermal noise.
A gap nobody could explain
For decades the spherical aberration of round lenses set the resolution of the electron microscope to about 100 electron wavelengths, the Scherzer resolution[3]. In the 15 years since aberration-corrected TEM was introduced[4], the resolving power improved to about 25 wavelengths at beam energies of 60 to 300 kV[5]–[9]. A better information limit drives atomic-resolution imaging in materials science[10],[11] and, at still lower beam energies, the study of light-atom materials that the beam would otherwise damage by knocking atoms out of place[12]–[14], among them the Ulm work on TEM at 20 kV.
The latest instruments corrected for both chromatic aberration Cc and spherical aberration Cs delivered convincing results. Their lateral incoherence and focus spread from residual aberrations and instabilities were shown to be so small that a clearly better information limit should have been reached than the one actually observed[2]. While developing new instruments, the CEOS team analyzed the discrepancy and ruled out electronics noise, AC stray fields, higher-order aberrations, Coulomb interaction within the beam and imperfect vacuum, but the mismatch remained. A similar gap appeared in aberration-corrected Lorentz microscopy, in which the specimen sits in a field-free region[15].
Careful measurements of the contrast transfer as a function of spatial frequency g showed that the missing contrast followed an isotropic envelope function exp[−2(πσg)2], with σ a characteristic image spread[2]: as if every point of the image were shifted by a small random amount. The resulting information limit, at a spatial frequency of about 1/(πσ), did not depend on the numerical aperture and scaled with the electron wavelength for different beam energies, the way a magnetic force would. Decoherence in the TEM had long been studied, from inelastic scattering in the specimen to aloof excitations of passing electrons[16],[17], because calculated and observed image contrast often disagreed[18]–[20]. None of these effects was strong enough to explain the loss of contrast seen at CEOS.
Noise from warm metal
What had not been considered was magnetic field noise from currents that heat alone drives through the conducting parts of the instrument: the lenses, the multipoles and the vacuum tubes the beam always travels through. In electronics, the thermal motion of conduction electrons produces a random voltage across every resistor, the Johnson–Nyquist noise[21],[22]; the same thermal currents also produce fluctuating magnetic fields[23]–[25]. The team realized that a Cc/Cs-corrected TEM must be especially sensitive to them, for two reasons: the corrector removes the main sources of incoherence, and its extra optics lengthen the beam path considerably. By their estimate, transverse fields of about 0.1 nT, hundreds of thousands of times weaker than the Earth’s magnetic field, acting over a few centimeters along the optic axis would already explain the observations.
Thermal noise has a fingerprint: it depends on temperature. “If Johnson noise is the primary reason, it should be possible to show a temperature dependence of the amount of image spread due to the thermodynamic nature of its origin,” the authors write. This motivated a decisive experiment.
A column with a cold test tube
To keep the situation simple, the team removed the aberration corrector from the microscope and put a cylindrical vacuum chamber, 60 mm wide, in its place. Inside it, a copper tube cooled with liquid nitrogen held exchangeable test tubes up to 10 mm in diameter and 50 cm long; a permalloy screen around the whole set-up shielded it from outside fields (Fig. 1). Two test tubes were compared, a stainless steel liner tube with an outer diameter of 3.0 mm and a stack of permalloy tubes, a magnetic nickel–iron alloy, 7.0 mm across, both with a bore of 2.7 mm. Extra lenses and deflectors set up two beam paths: in path A the objective aperture is demagnified right behind the objective lens, which makes the beam least sensitive to noise; in path B a parallel beam runs through the test tube, magnified by a factor of 2.4.
To measure the image spread alone, the team imaged a thin amorphous film of about 2 nm tungsten on carbon with strongly tilted illumination. In the diffractograms of such images, the Fourier transforms, bright achromatic rings appear[26]. Along these rings the contrast transfer is not affected by the temporal and lateral incoherence of the electron source, so the way the intensity fades along a ring measures the image spread directly.
Cold tubes, sharper images
The diffractograms tell the story at a glance (Fig. 2). Tilted illumination turns the contrast transfer into bright achromatic rings; how far out they reach before fading into the background marks the finest spacing the microscope still transfers. All images were taken at the same magnification, within 3 %, and with the same Nyquist frequency of 33/nm, so the rings of all set-ups can be compared directly. In the reference set-up, beam path A, the rings reached out to a spatial frequency of at least 14/nm, that is, spacings of 71 pm. With the parallel beam of path B and an empty cooler, the transfer ended at about 9/nm (111 pm). With a test tube in place at room temperature, the noise was much stronger: the information transfer ended at about 5/nm (200 pm) with stainless steel and at 3.8/nm (263 pm) with permalloy. Cooled to an average of about 120 K, both tubes transferred clearly finer detail. Without a test tube, cooling changed hardly anything, because the copper cooler is so wide. “This experiment clearly proves the thermodynamic nature of the observed noise effect and shows that the phenomenon is related to the material of the test tubes,” the authors conclude.
Putting a number on the blur
To quantify the effect, the team scanned the diffractograms along the achromatic rings and multiplied the scan of the reference experiment by the envelope exp[−2(πσg)2], with the image spread σ as the only free parameter, until it matched the scan with the test tube; in this way the properties of the amorphous sample drop out. For the cold permalloy tube the best fit gave σ = 67 pm, while 57 pm and 77 pm clearly missed the data (Fig. 3a); a single fit was accurate to about ±3 pm.
The image spread follows from the random deflection of the electrons. The noise fields change at kilohertz to megahertz frequencies, so slowly that each electron flies through a frozen field pattern, but every one of the roughly 109 electrons forming an image meets a different one, much as in the frozen-phonon picture of thermal vibrations[27]. The deflections average to zero, but their spread grows with the length of the tube, about 0.5 m, and with the distance over which the noise fields stay correlated, about one and a half times the tube radius. From σ the team derived the variance of the magnetic field noise, plotted against the temperature of the tube for both materials (Fig. 3b). At room temperature the electrons were deflected by 25 nrad on average in the permalloy tube and by 17 nrad in the stainless steel tube. In general, the noise variance scales as C·kBT/R2, with the tube radius R and a constant C that depends on the geometry and only weakly on the material[23],[25],[28],[29].
Two features of this law matter for instrument builders. First, the noise variance falls with the square of the tube radius: the copper cooler, 15 mm wide and thus more than five times wider than the 2.7 mm bore of the test tubes, adds about thirty times less noise, which is why cooling it alone changed hardly anything. In a corrector, by contrast, the beam runs through multipoles and liner tubes only a few millimetres wide. Second, the optics turn the deflection angles θ into a blur in the image: the image spread grows with the focal length f of the objective lens and with the aperture magnification M, σ2 = M2f2⟨θ2⟩. With M = 2.4 and f = 1.73 mm the set-up was made deliberately sensitive, so that a few tens of nanoradians showed up as a blur of tens of picometres. The same relation lets the measured noise be scaled to any other column, which is how the authors arrive at the 15 to 25 pm of real Cc/Cs-corrected microscopes.
How closely these results follow the theory of thermal noise is shown in the box “Validation: experiment vs. Johnson-noise theory” below, which also holds the test of isotropy (Fig. 4).
Microscope, test set-up and sample are listed in the box below.
Challenges named by the authors
Magnetic materials remain a problem: although the skin effect pushes their noise currents to the surface and lets the noise fade from frequencies as low as 1 to 10 Hz, they still add a large total variance, and the relocation of the currents raises the constant C by a factor of 2 to 4. Since the integrated noise hardly depends on the resistivity of a material over orders of magnitude, a better choice of metal is not an easy way out; the authors leave open whether a low-noise material with significant conductivity exists. The correlation length of the noise fields was stated without proof, and the temperature of the cold tube was known only within a range, which sets the horizontal error bars. The same limit, they note, also widens the effective source size in the scanning transmission electron microscope.
Outlook
Thermal magnetic field noise turned out to be a fundamental limit of electron microscopy that had gone unnoticed: a stochastic beam deflection by fields from the warm metal around the beam. Scaled to the Cc/Cs-corrected microscopes of the day, it readily explains the reported image spread of about 15 to 25 pm for a high-resolution objective lens[2]. For instrument builders it turns the bore of the multipoles, the lens gaps and the width of the liner tube into design parameters for coherence. “A detailed understanding of the origin and the properties of the thermal field noise should generally be considered very valuable for the design of future advanced instrumentation in charged particle optics,” the authors conclude. The link between thermodynamic current fluctuations and the optical transfer function of the microscope may also help to study magnetic field noise in other fields of science.
Resource:
Uhlemann, S., Müller, H., Hartel, P., Zach, J., & Haider, M. (2013).
Thermal Magnetic Field Noise Limits Resolution in Transmission Electron Microscopy.
Phys. Rev. Lett. 111(4), 046101.
https://doi.org/10.1103/PhysRevLett.111.046101
Update – August 4, 2014
The news article of July 22, 2013 ends above. Later work that builds on it is listed below.
A shorter corrector at Berkeley
Christian Kisielowski, Petra Specht and colleagues at the Lawrence Berkeley National Laboratory took up the result in Micron when they listed what an instrument needs to follow single atoms in real time[31]. In their reading, the CEOS study shows that resolution and stability are a trade-off against the hardware of aberration correction: a Cs corrector needs 30 to 40 additional power supplies, a Cc/Cs corrector more than 150, and it adds about half a meter to the column. They cite the image spread of 15 to 25 pm as “a hard limit on the resolution” of Cc/Cs-corrected systems such as their own TEAM I, a limit that scales with the length of the corrector. “Consequently, a shorter corrector allows for higher resolution and better stability,” they write. Their TEAM 0.5, with a Cs corrector and a monochromator only, showed no chromatic damping in a tilt series at 80 kV and reached 45 pm, which in their view supports the CEOS explanation.
Update – November 29, 2014
Theory meets the steel tube
In Ultramicroscopy, Uhlemann, Müller, Zach and Haider followed up with the theory behind their experiment[32]. From the fluctuation–dissipation theorem they derived semi-analytical models for the strength, bandwidth and spatial correlation of the noise fields, for a thin metal tube and for layers of magnetic and non-magnetic material. For the thin stainless steel tube the prediction now works without a free parameter: the measured slope of the noise against temperature gave a constant C of 0.193 ± 0.014, the theory 0.166 ± 0.005. The model also explains why a magnetic tube of the same size produces about twice the noise variance. A new measurement on a “tube-in-tube”, the stainless steel tube inside permalloy tubes, the arrangement found in most lenses and multipoles, behaved almost like the bare permalloy tube, again as the layered model predicts. The paper also recalls how the team got on the track: scaled from the tube experiments to the dimensions of their corrector, the noise predicted an image spread of πσ = 76 pm at 80 kV, close to the 80 pm they had measured before the cause was known. Even without a corrector, they estimate, the narrow bore of an objective pole piece can limit the transfer at 80 kV to about 30 pm. The authors doubt that much quieter materials exist, “but surprises would be very welcome of course.”
Update – August 9, 2016
SALVE built around the noise
Three years later the CEOS team, FEI and Ulm University presented the Cc/Cs corrector of the SALVE microscope (Linck et al.)[33]. Its designers had to take the thermal magnetic field noise into account from the start: the bore of the multipoles, the lens gaps and the diameter of the liner tube were balanced so that the corrector reached the required correction strength and focusing power without exposing the beam to too much image spread, “at the edge of technological feasibility,” with the magnetic materials in the multipoles used close to saturation. The SALVE microscope reached information limits from 76 pm at 80 kV to 139 pm at 20 kV (Table 2).
| Voltage | Wavelength | Information limit | Aperture equivalent |
|---|---|---|---|
| 80 kV | 4.176 pm | 76 pm | 55 mrad |
| 60 kV | 4.866 pm | 83 pm | 59 mrad |
| 40 kV | 6.016 pm | 90 pm | 67 mrad |
| 30 kV | 6.979 pm | 115 pm | 61 mrad |
| 20 kV | 8.589 pm | 139 pm | 62 mrad |
Table 2: Information limits of the Cc/Cs-corrected SALVE microscope measured with Young’s fringes[33].
Update – May 21, 2018
Image spread in the SALVE fit
Felix Börrnert, Julian Renner and Ute Kaiser at Ulm University measured the source brightness and the illumination semi-angle of the SALVE microscope by fitting the information limit against defocus (Börrnert et al.)[34]. In a chromatically corrected instrument such a fit works only if the image spread from magnetic noise enters it as one of the damping terms; for SALVE at 80 kV they used an image spread of 21 pm.
Update – August 30, 2018
Johnson noise sets the SALVE limit
In a detailed study of imaging at 80 and 20 kV, Börrnert and Kaiser found that the information limit of the SALVE microscope is no longer set by focus spread but is dominated by the Johnson–Nyquist noise and the image spread it causes (Börrnert et al.)[35]. From the finest nonlinearly transferred signals in images of graphene they derived an image spread of at most 22.5 pm at 80 kV, within the range of 15 to 25 pm that the CEOS study had attributed to thermal field noise. The limit even has an advantage: image spread damps nonlinear contrast transfer in the same way as linear transfer, so nearly all the information in a SALVE image lies within the information limit.
Update – March 29, 2023
Johnson noise in a laser phase plate
Ten years after the CEOS study the effect turned up in cryo-electron microscopy of proteins. Jeremy Axelrod, Holger Müller, Robert Glaeser and colleagues at the University of California, Berkeley, and the Lawrence Berkeley National Laboratory, together with Thermo Fisher Scientific, had built a laser phase plate, in which a laser beam focused in an optical cavity shifts the phase of the unscattered electrons[36]. To make room for it, extra relay optics magnify the diffraction pattern, and the electron beam had to pass a 23 mm long hole only 2 mm wide. Their first density map of a protein stopped at a modest 3.8 Å. Citing the work from CEOS, they identified thermal magnetic field noise from the conductive parts of the phase plate as the cause: with a 2 mm hole the contrast envelope near 3 Å fell almost by half, while an 8 mm hole restored it to nearly the value without any phase plate, as the Johnson-noise model predicts. What remains is noise from the liner tube of the column, together with the chromatic aberration of the relay optics. The next design widens the liner tube and makes the focal length of the relay optics adjustable; cooling the liner tube with liquid nitrogen, as in the CEOS experiment, is named as another option.
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Reich, E. S. (2013). Imaging Hits Noise Barrier. Nature, 499, 135–136. https://doi.org/10.1038/499135a
-
Haider, M., Hartel, P., Müller, H., Uhlemann, S., Zach, J. (2010). Information Transfer in a TEM Corrected for Spherical and Chromatic Aberration. Microscopy and Microanalysis, 16, 393–408. https://doi.org/10.1017/S1431927610013498
-
Scherzer, O. (1949). The Theoretical Resolution Limit of the Electron Microscope. Journal of Applied Physics, 20, 20–29. https://doi.org/10.1063/1.1698233
-
Haider, M., Uhlemann, S., Schwan, E., Rose, H., Kabius, B., Urban, K. (1998). Electron Microscopy Image Enhanced. Nature, 392, 768–769. https://doi.org/10.1038/33823
-
Urban, K. W. (2008). Studying Atomic Structures by Aberration-Corrected Transmission Electron Microscopy. Science, 321, 506–510. https://doi.org/10.1126/science.1152800
-
Kisielowski, C., Freitag, B., Bischoff, M., van Lin, H., Lazar, S., Knippels, G., Tiemeijer, P., van der Stam, M., von Harrach, S., Stekelenburg, M., et al. (2008). Detection of Single Atoms and Buried Defects in Three Dimensions by Aberration-Corrected Electron Microscope With 0.5-Å Information Limit. Microscopy and Microanalysis, 14, 469–477. https://doi.org/10.1017/S1431927608080902
-
Erni, R., Rossell, M. D., Kisielowski, C., Dahmen, U. (2009). Atomic-Resolution Imaging With a Sub-50-Pm Electron Probe. Physical Review Letters, 102, 096101. https://doi.org/10.1103/PhysRevLett.102.096101
-
Krivanek, O. L., Chisholm, M. F., Nicolosi, V., Pennycook, T. J., Corbin, G. J., Dellby, N., Murfitt, M. F., Own, C. S., Szilagyi, Z. S., Oxley, M. P., et al. (2010). Atom-by-Atom Structural and Chemical Analysis by Annular Dark-Field Electron Microscopy. Nature, 464, 571–574. https://doi.org/10.1038/nature08879
-
Batson, P. E., Dellby, N., Krivanek, O. L. (2002). Sub-ångstrom Resolution Using Aberration Corrected Electron Optics. Nature, 418, 617–620. https://doi.org/10.1038/nature00972
-
Urban, K. W. (2009). Is Science Prepared for Atomic-Resolution Electron Microscopy? Nature Materials, 8, 260–262. https://doi.org/10.1038/nmat2407
-
Segal, M. (2009). Surely You're Happy, Mr Feynman! Nature Nanotechnology, 4, 786–788. https://doi.org/10.1038/nnano.2009.360
-
Alem, N., Yazyev, O. V., Kisielowski, C., Denes, P., Dahmen, U., Hartel, P., Haider, M., Bischoff, M., Jiang, B., Louie, S. G., et al. (2011). Probing the Out-of-Plane Distortion of Single Point Defects in Atomically Thin Hexagonal Boron Nitride at the Picometer Scale. Physical Review Letters, 106, 126102. https://doi.org/10.1103/PhysRevLett.106.126102
-
Girit, Ç. Ö., Meyer, J. C., Erni, R., Rossell, M. D., Kisielowski, C., Yang, L., Park, C.-H., Crommie, M. F., Cohen, M. L., Louie, S. G., et al. (2009). Graphene at the Edge: Stability and Dynamics. Science, 323, 1705–1708. https://doi.org/10.1126/science.1166999
-
Kaiser, U., Biskupek, J., Meyer, J., Leschner, J., Lechner, L., Rose, H., Stöger-Pollach, M., Khlobystov, A., Hartel, P., Müller, H., et al. (2011). Transmission Electron Microscopy at 20kV for Imaging and Spectroscopy. Ultramicroscopy, 111, 1239–1246. https://doi.org/10.1016/j.ultramic.2011.03.012
-
Freitag, B., Bischoff, M., Mueller, H., Hartel, P., von Harrach, H. S. (2009). Sub-Nanometer Resolution in Field-Free Imaging Using a Titan 80-300 With Lorentz Lens and Image Cs-Corrector at 300kV Acceleration Voltage. Microscopy and Microanalysis, 15, 184–185. https://doi.org/10.1017/S143192760909429X
-
Sonnentag, P., Hasselbach, F. (2007). Measurement of Decoherence of Electron Waves and Visualization of the Quantum-Classical Transition. Physical Review Letters, 98, 200402. https://doi.org/10.1103/PhysRevLett.98.200402
-
Howie, A. (2011). Mechanisms of Decoherence in Electron Microscopy. Ultramicroscopy, 111, 761–767. https://doi.org/10.1016/j.ultramic.2010.07.008
-
Hÿtch, M. J., Stobbs, W. (1994). Quantitative Comparison of High Resolution TEM Images With Image Simulations. Ultramicroscopy, 53, 191–203. https://doi.org/10.1016/0304-3991(94)90034-5
-
Boothroyd, C. B. (2000). Quantification of High-Resolution Electron Microscope Images of Amorphous Carbon. Ultramicroscopy, 83, 159–168. https://doi.org/10.1016/S0304-3991(00)00012-7
-
Howie, A. (2004). Hunting the Stobbs Factor. Ultramicroscopy, 98, 73–79. https://doi.org/10.1016/j.ultramic.2003.08.002
-
Johnson, J. B. (1928). Thermal Agitation of Electricity in Conductors. Physical Review, 32, 97–109. https://doi.org/10.1103/PhysRev.32.97
-
Nyquist, H. (1928). Thermal Agitation of Electric Charge in Conductors. Physical Review, 32, 110–113. https://doi.org/10.1103/PhysRev.32.110
-
Roth, B. J. (1998). Thermal Fluctuations of the Magnetic Field Over a Thin Conducting Plate. Journal of Applied Physics, 83, 635–638. https://doi.org/10.1063/1.366753
-
Kornack, T. W., Smullin, S. J., Lee, S.-K., Romalis, M. V. (2007). A Low-Noise Ferrite Magnetic Shield. Applied Physics Letters, 90, 223501. https://doi.org/10.1063/1.2737357
-
Lee, S.-K., Romalis, M. V. (2008). Calculation of Magnetic Field Noise From High-Permeability Magnetic Shields and Conducting Objects With Simple Geometry. Journal of Applied Physics, 103, 084904. https://doi.org/10.1063/1.2885711
-
Boothroyd, C. B., Dunin-Borkowski, R. E. (2006). In: N. Tanaka (ed.), Proceedings of the 16th International Microscopy Congress, Sapporo, Japan. Elsevier, New York.
-
Loane, R. F., Xu, P., Silcox, J. (1991). Thermal Vibrations in Convergent-Beam Electron Diffraction. Acta Crystallographica Section A Foundations of Crystallography, 47, 267–278. https://doi.org/10.1107/S0108767391000375
-
Varpula, T., Poutanen, T. (1984). Magnetic Field Fluctuations Arising From Thermal Motion of Electric Charge in Conductors. Journal of Applied Physics, 55, 4015–4021. https://doi.org/10.1063/1.332990
-
Clem, J. (1987). Johnson Noise From Normal Metal Near a Superconducting SQUID Gradiometer Circuit. IEEE Transactions on Magnetics, 23, 1093–1096. https://doi.org/10.1109/TMAG.1987.1065127
-
Marks, L. (2013). What Are the Resolution Limits in Electron Microscopes? Physics, 6, 82. https://doi.org/10.1103/Physics.6.82
-
Kisielowski, C., Specht, P., Gygax, S. M., Barton, B., Calderon, H. A., Kang, J. H., Cieslinski, R. (2014). Instrumental Requirements for the Detection of Electron Beam-Induced Object Excitations at the Single Atom Level in High-Resolution Transmission Electron Microscopy. Micron, 68, 186–193. https://doi.org/10.1016/j.micron.2014.07.010
-
Uhlemann, S., Müller, H., Zach, J., Haider, M. (2015). Thermal Magnetic Field Noise: Electron Optics and Decoherence. Ultramicroscopy, 151, 199–210. https://doi.org/10.1016/j.ultramic.2014.11.022
-
Linck, M., Hartel, P., Uhlemann, S., Kahl, F., Müller, H., Zach, J., Haider, M., Niestadt, M., Bischoff, M., Biskupek, J., et al. (2016). Chromatic Aberration Correction for Atomic Resolution TEM Imaging From 20 to 80 kV. Phys. Rev. Lett., 117, 076101. https://doi.org/10.1103/PhysRevLett.117.076101
-
Börrnert, F., Renner, J., Kaiser, U. (2018). Electron Source Brightness and Illumination Semi-Angle Distribution Measurement in a Transmission Electron Microscope. Microsc. Microanal., 24, 249–255. https://doi.org/10.1017/S1431927618000223
-
Börrnert, F., Kaiser, U. (2018). Chromatic- and Geometric-Aberration-Corrected TEM Imaging at 80 kV and 20 kV. Phys. Rev. A, 98, 023861. https://doi.org/10.1103/PhysRevA.98.023861
-
Axelrod, J. J., Petrov, P. N., Zhang, J. T., Remis, J., Buijsse, B., Glaeser, R. M., Müller, H. (2023). Overcoming Resolution Loss Due to Thermal Magnetic Field Fluctuations From Phase Plates in Transmission Electron Microscopy. Ultramicroscopy, 249, 113730. https://doi.org/10.1016/j.ultramic.2023.113730