Residual Aberrations in HRTEM
March 18, 2012 – Even a well-tuned aberration corrector leaves small errors that drift back during the day. Researchers from Ulm and CEOS in Heidelberg set coma and two-fold astigmatism on purpose in an image-side corrected FEI Titan at 80 kV and imaged carbon nanotubes. Coma breaks the mirror symmetry of the two tube walls, a sign the usual power spectrum does not show; it can be seen in the image from about 50 nm on, and even on a microscope without corrector. Image simulations reproduce every setting, and drift measurements on two microscopes in Jülich confirm the time scales.
An aberration corrector removes the spherical aberration of the objective lens, but never all of its errors, and the ones left over drift back during the day. Researchers from Ulm University and CEOS in Heidelberg have tested what the two most common leftovers, coma and two-fold astigmatism, do to the image of a single-walled carbon nanotube. With an FEI Titan at 80 kV and its CEOS image corrector, they set each aberration on purpose and compared the images with simulations. Coma has a fingerprint that the usual power spectrum does not show: the two walls of the tube lose their mirror symmetry. An experienced user can see it directly in the image above about 50 nm of coma, and the same test works on a microscope without corrector.
Leftovers of the correction
Since aberration correctors became a reality[1], transmission electron microscopy has imaged light atoms that were out of reach before. Materials made of light atoms, such as graphene, nanotubes and polymers, are quickly damaged at 200 or 300 kV, so the voltage is lowered: the threshold for knocking a carbon atom out of a nanotube lies at 86 keV[2]. At 80 kV, however, the resolution is limited by the electron wavelength and by the energy spread of the beam. Without correction, the point resolution of a lens with a spherical aberration Cs of about 1 mm is given by Scherzer[3], about 200 pm at 300 kV:
With a corrector this limit moves into the sub-Ångström range, and the information limit set by the chromatic aberration Cc and the energy spread ΔE of the electron gun takes over[4]:
For a Schottky gun with 0.7 eV and Cc = 1.5 mm this gives 190 pm at 80 kV, more than the 150 pm length of a carbon bond. Lowering the energy spread of the gun[5] brings it down to 154 pm, enough for the carbon lattice of nanotubes and graphene, and Ulm had used these conditions for chemistry inside nanotubes[6],[7] (see the news articles Reactions in a Nano-Container and Sulphur-Terminated Nanoribbons[8],[9]).
Once Cs is gone, the smaller aberrations count: two-fold astigmatism A1, coma B2, three-fold astigmatism A2 and those of third order. They can be measured from focus or tilt series of images, and an exit wave reconstruction can remove them afterwards[10]; a way to judge them from a single image of a crystal had been proposed in theory[11] but not tried. During normal work, residual aberrations are always present, because the corrector drifts. How this happens and how operators deal with it is described in the box below.
A nanotube as test object
Silicon dumbbells along [110] are the classic test of resolution, but a dumbbell that looks uneven can have many causes: mistilt, a bent or thick sample, dynamic scattering or residual aberrations. Carbon nanotubes avoid this. They are thin and well defined, two atomic layers in projection, so neither the exact orientation nor dynamic effects matter at 80 kV[13]. The team used three tubes, a zigzag (19,0) and two armchair tubes, (10,10) and (11,11) (Fig. 1); their indices were read from the power spectra[14], and the results hold for any type of tube. The images were taken on the image-side corrected FEI Titan in Ulm at 80 kV, with the extraction voltage of the gun lowered to reduce the energy spread to 0.45 eV. The corrector was tuned until the residual second- and third-order aberrations were below the accuracy of the measurement, with Cs set slightly positive at (5 ± 1) µm. Then coma and astigmatism were set by hand with the calibrated controls of the corrector software, for example 500 nm of coma pointing parallel or perpendicular to the tube. Every setting was simulated with the multislice program QSTEM[15]. The details are listed in the box below.
The broken mirror
Fig. 2 shows what happens to the (19,0) tube. Without coma and astigmatism, its lattice is visible at Scherzer focus. With 500 nm of coma pointing along the tube, the lattice is blurred. With the same coma pointing across the tube, the lattice only loses contrast, but something else changes: the two walls no longer look alike. The upper wall turns from bright outside to dark inside, as in normal underfocus with positive Cs; the lower wall turns from dark outside to bright inside, as in overfocus. This loss of mirror symmetry is the typical sign of coma. Two-fold astigmatism behaves differently: 20 nm of A1 wipes out almost all lattice contrast, but the walls stay mirror-symmetric.
The (10,10) tube confirms it (Fig. 3). Coma of 660 nm across the tube makes the walls asymmetric; at 1320 nm the effect is stronger, and white fringes appear inside and outside the tube. Turning the coma by 180° swaps the black and white contrast of the walls, and so does a change from underfocus to overfocus. In every case the simulations reproduce the images, as the box below shows.
How little coma shows
500 nm of coma is far more than the corrector drifts during a day. To find the smallest coma that can still be seen, the team simulated the (10,10) tube with coma from 0 to 660 nm and the same noise as in the experiment, a dose of 6 × 106 e/nm2, and took line scans across the tube (Fig. 4). The height difference between the main maxima of the two walls measures the coma. At 50 nm, the difference is just below the noise and cannot be seen with the naked eye; from 100 nm on, it rises above the noise. The detection limit thus lies above 50 nm, slightly larger than the measuring accuracy of the corrector software at 80 kV, twice the π/4 limit for a single coefficient and half the limit when the image shift compensates part of the coma.
Coma without a corrector
Coma matters on a microscope without corrector too. With the hexapoles switched off, Cs was 1.3 mm, and coma of 1 µm still broke the mirror symmetry of an (11,11) tube, although the strong delocalization fringes came from Cs. The usual coma-free alignment by wobbling the beam tilt is only accurate to about 1 µm, so its error can be seen in the image of a tube. The box below compares the uncorrected and the corrected microscope (Fig. 5).
Challenges named by the authors
The authors name the limits of the method. Two-fold astigmatism can be corrected easily in the live power spectrum during recording, but coma cannot. Coma could be corrected in the live image when two crossing tubes allow its two directions to be judged from the symmetry of their walls, but such correction by hand is not practicable in most cases. Once the symmetry breaks, the authors recommend a classic Zemlin tableau, a set of power spectra at tilted beam, on thin carbon film nearby. At usual doses of about 106 e/nm2, coma below about 50 nm cannot be detected in the image. Free-hanging tubes vibrate and bend, which had to be added to the simulations by hand, and how fast the corrector drifts depends on the microscope and its room. How fast it drifts was measured at the same time on two microscopes in Jülich; the box below compares the findings.
A test in every image
“With the coma streak of B2 perpendicular to the elongation of the SWNTs the break-down of the mirror symmetry is obvious,” the authors conclude. “Amazingly this even holds true if Cs is not corrected and hence is the dominating aberration.” Every nanotube in the field of view is a built-in test of the corrector: if its walls look different, coma is back. The images were taken with the image-side aberration-corrected FEI Titan 80-300 at 80 kV at Ulm University.
Resource:
Biskupek, J., Hartel, P., Haider, M., & Kaiser, U. (2012).
Effects of residual aberrations explored on single-walled carbon nanotubes.
Ultramicroscopy 116, 1–7.
https://doi.org/10.1016/j.ultramic.2012.03.008
Update – October 15, 2012
The news article of March 18, 2012 ends above. Later work is listed below.
Why the corrected state does not last
Leiden University and the IBM T. J. Watson Research Center explained why the corrected state of a microscope lasts only a few minutes[19]. They cite the Ulm study, together with measurements in Jülich and Berkeley, as the experimental evidence. With their contrast transfer function near optimum correction, they showed that the corrected state is intrinsically unstable: the better the correction, the narrower the range of aberrations within which it holds, and the more strongly small mechanical and electronic drifts are amplified. To gain at least 90 % of the resolution that correcting Cs offers in a microscope also corrected for Cc, Cs must be corrected to within 1/10 000 of its uncorrected value and held there, about 0.1 µm for a lens with 1 mm. They propose an “instability budget” for trading resolution against stability, and control systems that readjust the microscope on the fly.
Update – April 1, 2015
Removing coma afterwards
Ulm and EPFL in Lausanne then removed the anti-symmetric residual aberrations, coma and three-fold astigmatism, from single images after recording[20]. For weakly scattering 2D materials such as graphene at 80 kV, the same aberrations with opposite sign are applied to the Fourier transform of the image. They start from the Ulm finding that residual aberrations are always present even in corrected microscopes, where A2 and B2 remain at about 100 nm, and the method needs no focal series. On the same image-side corrected FEI Titan at 80 kV, a defect in graphene whose lattice did not look hexagonal regained its hexagons after removing 150 nm of A2, and a mirror-twin boundary in MoSe2 looked the same on both sides after removing 75 nm. Even 1 µm of A2, set on purpose with the corrector, could be removed. The method also allows residual aberrations to be estimated by trial and error. The effect is not limited to HRTEM: in ADF-STEM of multilayer h-BN at 60 kV, Vienna and Paris-Saclay showed in 2026 that about 50 nm of A2 aligned with the lattice changes the brightness of neighbouring boron and nitrogen columns without visibly distorting them, so that defects and stacking can be misread[21].
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Haider, M., Rose, H., Uhlemann, S., Schwan, E., Kabius, B., Urban, K. (1998). A Spherical-Aberration-Corrected 200 kV Transmission Electron Microscope. Ultramicroscopy, 75, 53–60. https://doi.org/10.1016/S0304-3991(98)00048-5
-
Smith, B. W., Luzzi, D. E. (2001). Electron Irradiation Effects in Single Wall Carbon Nanotubes. J. Appl. Phys., 90, 3509–3515. https://doi.org/10.1063/1.1383020
-
Scherzer, O. (1949). The Theoretical Resolution Limit of the Electron Microscope. J. Appl. Phys., 20, 20–29. https://doi.org/10.1063/1.1698233
-
Haider, M., Hartel, P., Müller, H., Uhlemann, S., Zach, J. (2010). Information Transfer in a TEM Corrected for Spherical and Chromatic Aberration. Microsc. Microanal., 16, 393–408. https://doi.org/10.1017/S1431927610013498
-
Meyer, J. C., Chuvilin, A., Algara-Siller, G., Biskupek, J., Kaiser, U. (2009). Selective Sputtering and Atomic Resolution Imaging of Atomically Thin Boron Nitride Membranes. Nano Lett., 9, 2683–2689. https://doi.org/10.1021/nl9011497
-
Chuvilin, A., Khlobystov, A. N., Obergfell, D., Haluska, M., Yang, S., Roth, S., Kaiser, U. (2010). Observations of Chemical Reactions at the Atomic Scale: Dynamics of Metal‐Mediated Fullerene Coalescence and Nanotube Rupture. Angew. Chem. Int. Ed., 49, 193–196. https://doi.org/10.1002/anie.200902243
-
Chuvilin, A., Bichoutskaia, E., Gimenez-Lopez, M. C., Chamberlain, T. W., Rance, G. A., Kuganathan, N., Biskupek, J., Kaiser, U., Khlobystov, A. N. (2011). Self-Assembly of a Sulphur-Terminated Graphene Nanoribbon Within a Single-Walled Carbon Nanotube. Nat. Mater., 10, 687–692. https://doi.org/10.1038/nmat3082
-
SALVE news (2010). Reactions in a Nano-Container. salve-project.de, January 1, 2010. https://salve-project.de/newspress/viewing-a-chemical-reaction-inside-a-nano-container.html
-
SALVE news (2011). Sulphur-Terminated Nanoribbons. salve-project.de, August 7, 2011. https://salve-project.de/newspress/sulphur-terminated-graphene-nanoribbon.html
-
Meyer, R. R., Kirkland, A. I., Saxton, W. O. (2002). A New Method for the Determination of the Wave Aberration Function for High Resolution TEM. 1. Measurement of the symmetric aberrations. Ultramicroscopy, 92, 89–109. https://doi.org/10.1016/S0304-3991(02)00071-2
-
Stenkamp, D. (1998). Detection and Quantitative Assessment of Image Aberrations From Single HRTEM Lattice Images. J. Microsc., 190, 194–203. https://doi.org/10.1046/j.1365-2818.1998.3000848.x
-
Barthel, J., Thust, A. (2010). Aberration Measurement in HRTEM: Implementation and Diagnostic Use of Numerical Procedures for the Highly Precise Recognition of Diffractogram Patterns. Ultramicroscopy, 111, 27–46. https://doi.org/10.1016/j.ultramic.2010.09.007
-
Lee, Z., Meyer, J. C., Rose, H., Kaiser, U. (2012). Optimum HRTEM Image Contrast at 20 kV and 80 kV – Exemplified by Graphene. Ultramicroscopy, 112, 39–46. https://doi.org/10.1016/j.ultramic.2011.10.009
-
Qin, L.-C. (2007). Determination of the Chiral Indices (n,m) of Carbon Nanotubes by Electron Diffraction. Phys. Chem. Chem. Phys., 9, 31–48. https://doi.org/10.1039/B614121H
-
Koch, C. T. (2002). Determination of Core Structure Periodicity and Point Defect Density Along Dislocations. PhD thesis, Arizona State University.
-
Müller, H., Maßmann, I., Uhlemann, S., Hartel, P., Zach, J., Haider, M. (2011). Aplanatic Imaging Systems for the Transmission Electron Microscope. Nucl. Instrum. Methods Phys. Res. A, 645, 20–27. https://doi.org/10.1016/j.nima.2010.12.091
-
Rez, D., Rez, P., Grant, I. (1994). Dirac–Fock Calculations of X-Ray Scattering Factors and Contributions to the Mean Inner Potential for Electron Scattering. Acta Crystallogr. A, 50, 481–497. https://doi.org/10.1107/S0108767393013200
-
Leschner, J. (2011). High-Resolution Electron Tomography on Beam-Sensitive Carbon Materials. PhD thesis, Ulm University.
-
Schramm, S. M., van der Molen, S. J., Tromp, R. M. (2012). Intrinsic Instability of Aberration-Corrected Electron Microscopes. Phys. Rev. Lett., 109, 163901. https://doi.org/10.1103/PhysRevLett.109.163901
-
Lehtinen, O., Geiger, D., Lee, Z., Whitwick, M. B., Chen, M.-W., Kis, A., Kaiser, U. (2015). Numerical Correction of Anti-Symmetric Aberrations in Single HRTEM Images of Weakly Scattering 2D-Objects. Ultramicroscopy, 151, 130–135. https://doi.org/10.1016/j.ultramic.2014.09.010
-