Graphene as Substrate
April 1, 2011 – The first proposal of pristine graphene as a substrate for biological TEM: researchers from the Biozentrum of the University of Basel, the University of Texas at Austin and Ulm have transferred continuous single-layer graphene, grown on copper foil, onto standard TEM grids. Power spectra recorded with an aberration-corrected TEM at 80 kV and EELS show that pristine graphene adds up to ten times less background than amorphous carbon, close to that of an empty frame. On the new support, stained DNA is visible in strong contrast without metal shadowing.
Every molecule imaged in a transmission electron microscope lies on something. For decades that support has been a film of amorphous carbon a few nanometres thick, and its disordered atoms add their own grainy signal to every image. Researchers from the Biozentrum of the University of Basel, the group of Rodney S. Ruoff at the University of Texas at Austin and Ulm University have now put a single layer of pristine graphene, grown on copper, onto standard TEM grids. It is the first study to propose pristine graphene, grown by CVD and transferred onto standard holey grids, as a substrate for biological TEM. Power spectra at 80 kV and electron energy-loss spectroscopy (EELS) show that the new support adds almost nothing to the image: its background lies up to ten times below that of amorphous carbon and close to that of an empty frame.
Crystalline supports have been wanted since the 1960s, when cleaved graphite and vermiculite were tried, but they were too hard to prepare for routine use[1],[2]. A crystal is in effect transparent at resolutions below its lattice period, and above it the periodic signal can be subtracted[3]. The lattice of graphene repeats at 213 pm and 123 pm[4], finer than the details biological TEM usually resolves, and a single layer is only 0.34 nm thick[5], so little scattering is left to add noise. Graphene also survived the beam well: the threshold for knock-on damage of carbon lies at 86 keV[6], yet the substrates withstood up to 300 kV at the low doses of biological work, as little as 2000–3000 electrons per nm2.
In earlier work with the Max Planck Institute of Biochemistry in Martinsried and Ulm, Radosav Pantelic had introduced graphene oxide, a water-loving relative of graphene, as a support[7]. Its oxygen groups, however, add a weak background of their own, raise the thickness of a layer to about 1 nm[8] and, deposited from solution, leave overlapping platelets of uneven thickness. Chemical vapor deposition (CVD) on copper foil, developed in Austin, gives continuous graphene that is more than 95 % single layer[9], and later even completely single layer[10], on foils of any size.
From copper foil onto the grid
The graphene is moved straight from the copper foil onto gold Quantifoil grids: a solvent pulls the grid onto the graphene, and the copper is then etched away from below. The full recipe is in the box below.
Three supports side by side
The comparison uses power spectra: the Fourier transform of an image shows how much signal a support contributes at each spatial frequency. All three supports were imaged with an image-side aberration-corrected FEI Titan at 80 kV, 200 nm defocus and a pixel size of 87 pm, with the same nominal count of 104 electrons per pixel (Fig. 1).
Amorphous carbon, 3–4 nm thick, gives the familiar rings of the contrast transfer function, which reach out to 4.2 nm−1. Graphene oxide still shows these rings, with about 40 % less phase contrast, out to 3.5 nm−1. Pristine graphene shows no rings at all: its only phase contrast is the first lattice reflection at 213 pm, and this reflection is sharper than in graphene oxide, whose lattice the harsh oxidation leaves full of defects. Below its lattice period, a crystalline support contributes only a smooth background, mostly from inelastic scattering. For pristine graphene this background lies up to an order of magnitude below that of amorphous carbon, and only a little above the curve of an empty frame recorded with the same dose.
That the low background really comes from weak inelastic scattering was tested with a second, independent method, EELS, in the box below, which also holds the spectra (Fig. 2).
Clean enough after handling
In air, adsorbates gradually settle on the pristine graphene. After 90 min at room temperature, most of a grid is still near pristine, since the adsorbates first gather around the few defects in the lattice (Fig. 3). Even the areas worst covered by amorphous deposits still give less phase contrast than graphene oxide beyond 1.7 nm−1. Handling the grids within reasonable limits during sample preparation therefore keeps most of the benefit.
DNA without metal shadowing
As a practical test the researchers imaged DNA, which on amorphous carbon is so faint that it is usually shadowed with metal. A double-stranded plasmid of about 7000 base pairs, encoding the chloride–proton antiporter of Escherichia coli, was spread without proteins on a monolayer of the spreading agent DMP-30 at the surface of a droplet[13], picked up with a graphene grid baked at 400 °C, and stained for 1 min with 1 % uranyl acetate. Imaged in bright field with an FEI Titan Krios at 80 kV, cooled with liquid nitrogen, at 2.5 µm defocus and a pixel size of 440 pm, the relaxed circular plasmid stands out in strong contrast without any metal shadowing (Fig. 4).
The background in the image is not zero. Because the support is so transparent, buffer components now show that would vanish on a thicker film; most of the remaining grain is probably the bound DMP-30 layer.
Challenges named by the authors
Without oxygen groups, pristine graphene repels water, which limits it for now to hydrophobic molecules and monolayer-based methods. The usual glow discharge or plasma cleaning that makes supports hydrophilic does not work: the electrons and ions knock atoms out of the thin graphene and etch it. Wet etching of the copper leaves precipitates and contamination that must be filtered and washed away, and in air the clean surface slowly collects adsorbates. The authors propose a gentle in-situ treatment, a controlled chemical functionalization or annealing in gas, that makes graphene hydrophilic without spoiling its properties, and name it as the focus of their ongoing work.
“As a purely crystalline substrate free from attenuation by oxidization, pristine graphene is transparent to TEM,” the authors conclude. Unlike graphene oxide from solution, the transferred CVD graphene is thinner and homogeneous in thickness, and it may take over wherever amorphous carbon has so far been the only choice.
Resource:
Pantelic, R. S., Suk, J. W., Magnuson, C. W., Meyer, J. C., Wachsmuth, P., Kaiser, U., Ruoff, R. S., & Stahlberg, H. (2011).
Graphene: Substrate preparation and introduction.
J. Struct. Biol. 174(1), 234–238.
https://doi.org/10.1016/j.jsb.2010.10.002
Update – April 20, 2014
The news article of April 1, 2011 ends above. Results published after it that test its findings are listed below.
Graphene made hydrophilic
Three years later, the MRC Laboratory of Molecular Biology in Cambridge closed the last gap this article had left open. There, Christopher Russo and Lori Passmore found the treatment the authors had called for, and they cite this work among the first tests of graphene for biological samples[14]. They exposed CVD graphene on Quantifoil grids to a pure hydrogen plasma of very low energy. Its hydrogen species carry 10–15 eV, too little to knock a carbon atom out of its sp2 bonds (21 eV), so the plasma adds hydrogen to the sheet instead of sputtering it. With increasing plasma dose the contact angle of water dropped from 91° to a saturation value of 66°, and already one hydrogen atom for every 20 carbon atoms made the surface hydrophilic enough for proteins (Fig. 5a). Electron diffraction at 300 kV showed the sharp graphene reflections unchanged after 20 s and 40 s of plasma (Fig. 5b–d): the lattice survives[14],[15].
The plasma also answers the second problem named above, the adsorbates. A 30 s treatment removed most of the contamination left by the transfer (Fig. 5e, f), because hydrocarbons break down far faster than the graphene bonds. In the power spectral density of the images, the cleaned graphene then lay clearly below a 2.8 nm film of amorphous carbon and was featureless (Fig. 5g), the same result this article found with power spectra at 80 kV (Fig. 1)[14].
With the plasma dose, the number of proteins on the graphene can be set. On grids without graphene only about 60 ribosomes per µm2 stayed in the ice, after 20 s of plasma there were about 600 and after 40 s about 1900 (Fig. 6a). Proteasomes and apoferritin adsorbed in the same way (Fig. 6b, c); apoferritin needed about 150 times less protein than in earlier cryo-EM studies. A reconstruction of the 80S ribosome from 20,050 particles on graphene reached 5.0 Å. Movies from a direct electron detector showed that the particles move less on graphene during exposure: in the first, fast phase 10–20 % and in the slower second phase 30–50 % slower than on amorphous carbon, and the results varied far less from grid to grid. The B factor, a measure of image blurring, fell by about 35 %[14].
| Question | This work (2011) | Russo & Passmore (2014) | Agreement |
|---|---|---|---|
| Support | CVD graphene from copper, transferred to Quantifoil grids | CVD graphene from copper, transferred to Quantifoil grids | same route |
| Background in the image | power spectra at 80 kV: up to 10× below amorphous carbon, close to an empty frame | power spectral density at 300 kV: clearly below 2.8 nm amorphous carbon, featureless | confirmed |
| Graphene oxide | adds an amorphous part (rings, broader EELS loss) | nearly as much background as thin amorphous carbon, often insulating, weaker | confirmed: pristine graphene preferred |
| Adsorbates | collect in air, most of the grid clean after 90 min | removed by 30 s of hydrogen plasma | problem solved |
| Water repellence | open problem; glow discharge etches graphene | contact angle 91° → 66°, lattice intact, proteins adsorb | problem solved |
| Biological test | stained DNA without metal shadowing | frozen ribosomes, proteasomes, apoferritin; 80S at 5.0 Å | extended to cryo-EM |
Table 2: Pristine graphene as a TEM support in this work and in the later study from Cambridge[14].
With hydrogen plasma, the last open problem of 2011 was solved. The treatment turned the pristine graphene support first proposed in this article into a routine substrate for frozen-hydrated proteins in single-particle cryo-EM, from transfer of the CVD sheet onto holey grids to the low background in the image.
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Dobelle, W. H., Beer, M. (1968). Chemically Cleaved Graphite Support Films for Electron Microscopy. J. Cell Biol., 39, 733–735. https://doi.org/10.1083/jcb.39.3.733
-
Hahn, M., Baumeister, W. (1974). High Resolution Negative Staining of Ferritin Molecules on Vermiculite Single Crystal Supports. Biochim. Biophys. Acta, 371, 267–274. https://doi.org/10.1016/0005-2795(74)90176-7
-
Meyer, J. C., Girit, C. O., Crommie, M. F., Zettl, A. (2008). Imaging and Dynamics of Light Atoms and Molecules on Graphene. Nature, 454, 319–322. https://doi.org/10.1038/nature07094
-
Meyer, J. C., Geim, A. K., Katsnelson, M. I., Novoselov, K. S., Booth, T. J., Roth, S. (2007). The Structure of Suspended Graphene Sheets. Nature, 446, 60–63. https://doi.org/10.1038/nature05545
-
Eda, G., Fanchini, G., Chhowalla, M. (2008). Large-Area Ultrathin Films of Reduced Graphene Oxide as a Transparent and Flexible Electronic Material. Nat. Nanotechnol., 3, 270–274. https://doi.org/10.1038/nnano.2008.83
-
Zobelli, A., Gloter, A., Ewels, C. P., Seifert, G., Colliex, C. (2007). Electron Knock-on Cross Section of Carbon and Boron Nitride Nanotubes. Phys. Rev. B, 75, 245402. https://doi.org/10.1103/PhysRevB.75.245402
-
Pantelic, R. S., Meyer, J. C., Kaiser, U., Baumeister, W., Plitzko, J. M. (2010). Graphene Oxide: A Substrate for Optimizing Preparations of Frozen-Hydrated Samples. J. Struct. Biol., 170, 152–156. https://doi.org/10.1016/j.jsb.2009.12.020
-
Stankovich, S., Dikin, D. A., Dommett, G. H. B., Kohlhaas, K. M., Zimney, E. J., Stach, E. A., Piner, R. D., Nguyen, S. T., Ruoff, R. S. (2006). Graphene-Based Composite Materials. Nature, 442, 282–286. https://doi.org/10.1038/nature04969
-
Li, X., Cai, W., An, J., Kim, S., Nah, J., Yang, D., Piner, R., Velamakanni, A., Jung, I., Tutuc, E., Banerjee, S. K., Colombo, L., Ruoff, R. S. (2009). Large-Area Synthesis of High-Quality and Uniform Graphene Films on Copper Foils. Science, 324, 1312–1314. https://doi.org/10.1126/science.1171245
-
Li, X., Zhu, Y., Cai, W., Borysiak, M., Han, B., Chen, D., Piner, R. D., Colombo, L., Ruoff, R. S. (2009). Transfer of Large-Area Graphene Films for High-Performance Transparent Conductive Electrodes. Nano Lett., 9, 4359–4363. https://doi.org/10.1021/nl902623y
-
Meyer, J. C., Girit, C. O., Crommie, M. F., Zettl, A. (2008). Hydrocarbon Lithography on Graphene Membranes. Appl. Phys. Lett., 92, 123110. https://doi.org/10.1063/1.2901147
-
Paredes, J. I., Villar-Rodil, S., Martínez-Alonso, A., Tascón, J. M. D. (2008). Graphene Oxide Dispersions in Organic Solvents. Langmuir, 24, 10560–10564. https://doi.org/10.1021/la801744a
-
Bratosin-Guttman, S. (1992). A New Method for the Routine Spreading of DNA in Protein-Free Conditions. J. Struct. Biol., 108, 162–167. https://doi.org/10.1016/1047-8477(92)90015-3
-
Russo, C. J., Passmore, L. A. (2014). Controlling Protein Adsorption on Graphene for Cryo-EM Using Low-Energy Hydrogen Plasmas. Nat. Methods, 11, 649–652. https://doi.org/10.1038/nmeth.2931
-
Russo, C. J., Passmore, L. A. (2014). Erratum: Controlling Protein Adsorption on Graphene for Cryo-EM Using Low-Energy Hydrogen Plasmas. Nat. Methods, 11, 773. https://doi.org/10.1038/nmeth0714-773e