Optimum HRTEM Contrast
January 1, 2012 – How much contrast can a single layer of carbon atoms give in the electron microscope? Researchers at Ulm University calculated the optimum HRTEM contrast of graphene at 80 and 20 kV, summing images over the electron energy distribution measured by EELS. At 80 kV the calculation matched focal series from the FEI Titan in Ulm with about 8.5 % atom contrast and no Stobbs factor, and a monochromator would double it. At 20 kV, high contrast for light atoms needs a corrector for the chromatic aberration as well, the goal of the SALVE microscope.
How bright can a single layer of carbon atoms look in the electron microscope, and what does it take to make it brighter? Researchers at Ulm University worked out the optimum image contrast of graphene for an aberration-corrected microscope at 80 kV and at 20 kV. Instead of the usual approximation for the energy spread of the electrons, they summed images over the energy distribution actually measured by electron energy loss spectroscopy (EELS). At 80 kV the calculation matched the experiment on the FEI Titan in Ulm without any correction factor: about 8.5 % atom contrast, the same for bright and dark atoms. A monochromator would double it. At 20 kV, however, the contrast barely rises unless the chromatic aberration is corrected as well; then it reaches more than 40 %. The study gave the theoretical reason for the Cc/Cs-corrected microscope that the SALVE project was building.
Why fewer kilovolts
Hardware correction of the spherical aberration Cs brought atomic resolution to transmission electron microscopy (TEM)[1],[2], and with it the old dream of imaging sensitive organic structures with light atoms came closer, thanks to a new generation of low-voltage microscopes. Graphene, a single layer of carbon atoms, is the thinnest sample there is and can be prepared as a clean, free-standing membrane[3]. An 80 kV beam is usually gentle enough for perfect graphene, but carbon atoms at defects and edges are knocked out even below 80 kV[4], which makes still lower voltages desirable. Ulm had just recorded the first Cs-corrected images of graphene and silicon at 20 kV[5] (see TEM at 20 kV[6]), and within the SALVE project a microscope was under construction that corrects both the geometric and the chromatic aberrations of the objective lens[7], for a resolution of 172 pm at 20 kV. This prompted the team to look again at how the contrast of such images is modelled.
Every electron energy counts
Image calculations usually describe the limited coherence of the beam with envelope functions that damp the contrast at high spatial frequencies. For a sample as thin as graphene, it is tempting to go one step further and treat it as a weak phase object, with a linear imaging theory on top. The authors set out to test whether such approximations hold for graphene. They used an approach in which many images, each for a slightly different electron energy, are added up[8],[9]. As weights they took the zero-loss peak of EELS spectra recorded on single-layer graphene, which contains the energy spread of the source together with all instabilities of the microscope. The projected potential of the carbon atoms came from density functional theory (DFT)[10] (see Charges Under the Microscope[11]); with it, the graphene contrast is about 8 % lower than with independent atoms. For every pair of Cs and defocus Δf, the program computed the image and its atom contrast, and the pair with the highest contrast was selected. The four steps and the formulas are given in the box below.
Bright atoms, dark atoms
Fig. 1 shows a section of the resulting table of calculated images for 80 kV. Surprisingly, the images look very similar over a wide range of Cs and Δf. Along diagonals of the table the contrast stays the same, and the highest values for dark and for bright atoms lie on the middle diagonals: −8.8 % for bright atoms at Cs = 0 and Δf = +6 nm, and +8.3 % for dark atoms at Δf = −5 nm. For a microscope without chromatic aberration (Cc = 0), the same search gives −19.4 % and +15.8 % (Fig. 2a). The optimum images with the real Cc of 1.41 mm and the measured energy width of 0.35 eV match the experimental images (Fig. 2b) within the error margins. Within the error margins, the authors write, “there is no Stobbs factor between experimental and calculated images”. The Stobbs factor is the notorious gap of up to a factor of three between measured and calculated contrast in HRTEM. Even the conventional envelope approach, with a focal spread of 4 nm, gave a similar 9.0 %. Microscope and samples are listed in the box below.
Twenty kilovolts need a Cc corrector
At 20 kV the picture changes (Fig. 3). With the same corrector and an energy width of 0.12 eV, reached with a monochromator, the optimum contrast is only −4.7 % and +6.2 %, not more than at 80 kV. Only when the chromatic aberration is removed, by a Cc corrector or an ideal monochromator, does the contrast rise steeply, to −41.6 % for bright atoms and +28.1 % for dark atoms. Here bright and dark contrast differ strongly, while at 80 kV they were nearly equal. This difference is the fingerprint of non-linear contributions: with longer wavelength the interaction with the sample grows, and at 20 kV even the light carbon atom is a strong scatterer. Even graphene, the authors write, “cannot be treated as a weak phase object”. A monochromator that removed all chromatic blurring would also remove an intolerably large share of the electrons. Fig. 4 shows how the contrast falls with Cc and with the energy width; the comparison of the two voltages is summarised in the box below.
The experiment at 80 kV
To find the highest contrast in the experiment, the team recorded a closely spaced focal series of single-layer graphene with Cs close to zero, in steps of 0.53 nm (Fig. 5). Each image covered only about 20 unit cells but used the full 2048 × 2048 pixels of the camera. The contrast was averaged over the clean unit cells. At a more common sampling of about 5 pixels per Å, the contrast would have been about half as large, because the camera blurs fine detail, as its measured modulation transfer function (MTF) shows[16]. The measured maxima were 8.5 % for bright atoms at Δf = +2.6 nm and 8.4 % for dark atoms at −3.7 nm, nearly identical, as the calculation predicted. At 80 kV, graphene can therefore be treated as a weak phase object with sufficient accuracy. The boxes below show how two other laboratories measured the same, with the graphene images from Dresden and Ulm at the same scale (Fig. 6), and how calculation and experiment compare in detail.
Challenges named by the authors
The authors name the limits of their study. Inelastic scattering, and with it phonon scattering, is not included; at 20 kV it may contribute significantly. No atomically resolved experimental images at 20 kV were available for comparison, because the corrector of the preliminary set-up compensated the geometric aberrations only up to 25 mrad, while graphene at 20 kV needs 40 mrad. A monochromator that removes the chromatic blurring completely would cost too much beam current. And in the experiment, the sign of the contrast is difficult to determine close to the contrast minima, so all measured values are given as positive numbers.
A case for chromatic correction
“Significant improvement of contrast can only be achieved by the additional correction of chromatic aberration,” the authors conclude for 20 kV. At 80 kV the new calculation agreed with both the conventional envelope method and the experiment, with neither a Stobbs factor nor a difference between bright and dark atoms, and a monochromator of about 0.1 eV would double the contrast. At 20 kV, all atoms are strong scatterers, and high contrast for light atoms needs a Cc corrector together with bright-atom imaging conditions. These were the design goals of the SALVE microscope. The experiments were done on the image-side aberration-corrected FEI Titan 80-300 at 80 kV and the monochromated Zeiss Libra 200 at Ulm University.
Resource:
Lee, Z., Meyer, J. C., Rose, H., Kaiser, U. (2012).
Optimum HRTEM image contrast at 20 kV and 80 kV – Exemplified by graphene.
Ultramicroscopy 112(1), 39–46.
https://doi.org/10.1016/j.ultramic.2011.10.009
Update – June 7, 2013
The news article of January 1, 2012 ends above. Later work is listed below.
Inelastic scattering at 20 kV
The same Ulm team took up the first challenge and added inelastic scattering to the calculation, for graphene and silicon at 20 kV[20]. They used EELS spectra from the SALVE I prototype, a Libra-based microscope at 20 kV, and simulated energy-filtered images for the Cc/Cs-corrected SALVE II. In graphene, less than 1 % of the electrons are scattered inelastically, yet the interference of these waves lowers the intensity of the zero-loss filtered image to 72 % of the purely elastic case. With Cc correction, a plasmon-loss filtered image keeps the elastic image contrast.
Update – February 6, 2014
Contrast at limited dose
In 2014 the team added the electron dose, which the 2012 study had taken as infinite[21] (see Improved evaluation of dose-dependent atom contrast based on image simulation for LVEM[22]). Built on the 2012 image model and on MTFs measured for the SALVE microscope at 20 to 80 kV, the calculations showed that signal-to-noise ratio, contrast and resolution all improve with dose. They confirmed the role of sampling found in 2012: for high atom contrast, a sampling of 7.5 pm per pixel or finer is needed.
Update – April 24, 2015
Large angles in STEM
JEOL Ltd. in Akishima and the AIST in Tsukuba went the other way, to scanning TEM at 30 kV[23]. Citing the Ulm study among the first lattice images of graphene at 20 kV, they used a corrector for higher-order aberrations that keeps the phase flat up to 64 mrad. Below 35 mrad, single carbon atoms on graphene could not be resolved; above 40 mrad, carbon atoms 142 pm apart were separated in dark-field images with a cold field emission gun. As the Ulm study had argued for HRTEM, large aperture angles are the key at low voltage.
Update – August 9, 2016
The Cc corrector at work
CEOS in Heidelberg, FEI in Eindhoven and Ulm University presented the Cc/Cs corrector of the SALVE microscope, which corrects chromatic and spherical aberration from 20 to 80 kV[24] (see SALVE Cc Correction Realized[25]). Over the whole voltage range it gives optimum conditions for bright-atom phase contrast within an aperture of at least 55 mrad, and the information transfer is no longer limited by chromatic aberration. At 30 kV, images of graphene showed reflections down to 107 pm and a strongly improved single-atom contrast in good agreement with calculated images. The combination the 2012 study had called for, Cc correction and bright atoms, was thus realised.
Update – February 10, 2019
Which image model to use
Finally, Ulm compared three ways of handling partial coherence in image simulation at 40 to 80 kV with experiments on MoTe2 from the FEI Titan and the SALVE microscope[26]: the incoherent summation used in 2012, the transmission cross-coefficient, and the faster envelope (quasi-coherent) model. The optimum experimental images were again selected by the contrast evaluation of the 2012 study. The envelope model suffices for weak phase objects and, with Cc correction, even for a few layers of heavy atoms. With Cs correction alone, heavy samples need the incoherent summation or the transmission cross-coefficient.
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Rose, H. (1990). Outline of a Spherically Corrected Semi-Aplanatic Medium-Voltage Transmission Electron Microscope. Optik, 85, 19–24.
-
Haider, M., Rose, H., Uhlemann, S., Kabius, B., Urban, K. (1998). Towards 0.1 Nm Resolution With the First Spherically Corrected Transmission Electron Microscope. J. Electron Microsc., 47, 395–405. https://doi.org/10.1093/oxfordjournals.jmicro.a023610
-
Meyer, J. C., Geim, A. K., Katsnelson, M. I., Novoselov, K. S., Booth, T. J., Roth, S. (2007). The Structure of Suspended Graphene Sheets. Nature, 446, 60–63. https://doi.org/10.1038/nature05545
-
Girit, C. O., Meyer, J. C., Erni, R., et al. (2009). Graphene at the Edge: Stability and Dynamics. Science, 323, 1705–1708. https://doi.org/10.1126/science.1166999
-
Kaiser, U., Biskupek, J., Meyer, J. C., et al. (2011). Transmission Electron Microscopy at 20 kV for Imaging and Spectroscopy. Ultramicroscopy, 111, 1239–1246. https://doi.org/10.1016/j.ultramic.2011.03.012
-
SALVE news (2011). TEM at 20 kV. salve-project.de, April 1, 2011. https://salve-project.de/newspress/transmission-electron-microscopy-at-20-kv.html
-
Haider, M., Müller, H., Uhlemann, S., Zach, J., Loebau, U., Hoeschen, R. (2008). Prerequisites for a Cc/Cs-Corrected Ultrahigh-Resolution TEM. Ultramicroscopy, 108, 167–178. https://doi.org/10.1016/j.ultramic.2007.07.007
-
Reimer, L., Kohl, H. (2008). Transmission Electron Microscopy: Physics of Image Formation. Springer, New York. https://doi.org/10.1007/978-0-387-40093-8
-
Rose, H. (2010). Theoretical Aspects of Image Formation in the Aberration-Corrected Electron Microscope. Ultramicroscopy, 110, 488–499. https://doi.org/10.1016/j.ultramic.2009.10.003
-
Meyer, J. C., Kurasch, S., Park, H. J., et al. (2011). Experimental Analysis of Charge Redistribution Due to Chemical Bonding by High-Resolution Transmission Electron Microscopy. Nat. Mater., 10, 209–215. https://doi.org/10.1038/nmat2941
-
SALVE news (2011). Charges Under the Microscope. salve-project.de, January 16, 2011. https://salve-project.de/newspress/charges_under_the_microscope.html
-
Watson, A. B., Barlow, H. B., Robson, J. G. (1983). What Does the Eye See Best?. Nature, 302, 419–422. https://doi.org/10.1038/302419a0
-
Zernike, F. (1942). Phase Contrast, a New Method for the Microscopic Observation of Transparent Objects. Physica, 9, 686–698. https://doi.org/10.1016/S0031-8914(42)80035-X
-
Lentzen, M. (2004). The Tuning of a Zernike Phase Plate With Defocus and Variable Spherical Aberration and Its Use in HRTEM Imaging. Ultramicroscopy, 99, 211–220. https://doi.org/10.1016/j.ultramic.2003.12.007
-
Meyer, J. C., Chuvilin, A., Algara-Siller, G., Biskupek, J., Kaiser, U. (2009). Selective Sputtering and Atomic Resolution Imaging of Atomically Thin Boron Nitride Membranes. Nano Lett., 9, 2683–2689. https://doi.org/10.1021/nl9011497
-
Thust, A. (2009). High-Resolution Transmission Electron Microscopy on an Absolute Contrast Scale. Phys. Rev. Lett., 102, 220801. https://doi.org/10.1103/PhysRevLett.102.220801
-
Meyer, J. C., Girit, C. O., Crommie, M. F., Zettl, A. (2008). Hydrocarbon Lithography on Graphene Membranes. Appl. Phys. Lett., 92, 123110. https://doi.org/10.1063/1.2901147
-
Jia, C. L., Houben, L., Thust, A., Barthel, J. (2010). On the Benefit of the Negative-Spherical-Aberration Imaging Technique for Quantitative HRTEM. Ultramicroscopy, 110, 500–505. https://doi.org/10.1016/j.ultramic.2009.10.006
-
Börrnert, F., Bachmatiuk, A., Gorantla, S., Wolf, D., Lubk, A., Büchner, B., Rümmeli, M. H. (2013). Retro-Fitting an Older (S)TEM With Two Cs Aberration Correctors for 80 kV and 60 kV Operation. J. Microsc., 249, 87–92 (available online 23 November 2012). https://doi.org/10.1111/j.1365-2818.2012.03684.x
-
Lee, Z., Rose, H., Hambach, R., Wachsmuth, P., Kaiser, U. (2013). The Influence of Inelastic Scattering on EFTEM Images – Exemplified at 20 kV for Graphene and Silicon. Ultramicroscopy, 134, 102–112. https://doi.org/10.1016/j.ultramic.2013.05.020
-
Lee, Z., Rose, H., Lehtinen, O., Biskupek, J., Kaiser, U. (2014). Electron Dose Dependence of Signal-to-Noise Ratio, Atom Contrast and Resolution in Transmission Electron Microscope Images. Ultramicroscopy, 145, 3–12. https://doi.org/10.1016/j.ultramic.2014.01.010
-
SALVE news (2014). Improved Evaluation of Dose-Dependent Atom Contrast Based on Image Simulation for LVEM. salve-project.de, October 12, 2014. https://salve-project.de/newspress/dose-dependent-image-simulation-for-tem.html
-
Sawada, H., Sasaki, T., Hosokawa, F., Suenaga, K. (2015). Atomic-Resolution STEM Imaging of Graphene at Low Voltage of 30 kV With Resolution Enhancement by Using Large Convergence Angle. Phys. Rev. Lett., 114, 166102. https://doi.org/10.1103/PhysRevLett.114.166102
-
Linck, M., Hartel, P., Uhlemann, S., Kahl, F., Müller, H., Zach, J., Haider, M., Niestadt, M., Bischoff, M., Biskupek, J., Lee, Z., Lehnert, T., Börrnert, F., Rose, H., Kaiser, U. (2016). Chromatic Aberration Correction for Atomic Resolution TEM Imaging From 20 to 80 kV. Phys. Rev. Lett., 117, 076101. https://doi.org/10.1103/PhysRevLett.117.076101
-
SALVE news (2016). SALVE Cc Correction Realized. salve-project.de, August 9, 2016. https://salve-project.de/newspress/SALVE-CC-correction-realized.html
-
Lee, Z., Lehnert, T., Kaiser, U., Rose, H. (2019). Comparison of Different Imaging Models Handling Partial Coherence for Aberration-Corrected HRTEM at 40–80 kV. Ultramicroscopy, 203, 68–75. https://doi.org/10.1016/j.ultramic.2019.01.007