Numerical Aberration Correction
April 1, 2015 – Even a well-tuned aberration corrector leaves residual aberrations of about 100 nm, and in 2D materials they can fake structures: a graphene lattice that is not hexagonal, or a MoSe2 crystal that looks like two different materials across a mirror-twin boundary. Researchers from Ulm University and the EPFL in Lausanne remove the anti-symmetric part of these aberrations, three-fold astigmatism and coma, from a single HRTEM image afterwards, with one Fourier transform and no focal series. Tested on simulations and on images from the FEI Titan at 80 kV, the method even removed 1 µm of three-fold astigmatism set on purpose.
A hardware aberration corrector makes the image of a microscope nearly free of aberrations, but only nearly: small residual aberrations always remain, and they can turn a clear lattice into a misleading pattern. Researchers from Ulm University and the EPFL in Lausanne showed that for thin 2D materials one group of these leftovers, the anti-symmetric aberrations such as three-fold astigmatism A2 and coma B2, can be removed from a single recorded image afterwards. No focal series and no hologram is needed, only the Fourier transform of the image and the aberrations with opposite sign. With the image-side corrected FEI Titan at 80 kV in Ulm, they restored the hexagonal lattice around a defect in graphene and made both sides of a mirror-twin boundary in MoSe2 look alike again. Even 1 µm of A2, set on purpose with the corrector, could be removed, and the grain boundary and the vacancy hidden under it came to light.
Nearly free of aberrations
Since hardware aberration correction became a reality, the point resolution of high-resolution transmission electron microscopy (HRTEM) is no longer limited by the spherical aberration Cs[1], and materials are imaged at atomic resolution even at low voltages[2],[3]. Once the strong Cs is removed, smaller aberrations such as three-fold astigmatism A2 and coma B2 become measurable and can be corrected[4], but only down to the accuracy with which they are measured and with which the corrector elements are set. “In the context of this paper, the word nearly should be emphasized,” the authors write about today’s nearly aberration-free images. Residual aberrations are always present, even in corrected instruments[5] (see the news article Residual Aberrations in HRTEM[6]), and the microscope drifts away from the corrected state as time passes after tuning[7]. Some residual aberrations are even wanted: a small positive Cs is needed for good phase contrast[4].
When the lattice misleads
A2 and B2 can remain in the order of 100 nm after correction, and Fig. 1 shows what this does[8]. The first image shows a point defect in single-layer graphene, produced by depositing carbon atoms on the sheet[9] (see First Observation of Self-Interstitials in Graphene[10]). Around it the graphene lattice does not look hexagonal. The cause is A2, which has three axes of symmetry in the image plane: depending on how it is oriented to the lattice, one of the two sublattices of graphene gives more contrast than the other. The structure of the defect can still be worked out, but the image is far from optimal.
The second image shows mirror-twin boundaries in single-layer MoSe2 (see Se-Deficient 2D MoSe2[11]). In an image without aberrations the Se sites are brighter than the Mo sites, because two Se atoms (Z = 34) always sit on top of each other, against a single Mo atom (Z = 42). In the recorded image, however, the lattice looks completely different on the two sides of each boundary. On one side the second sublattice is much brighter; on the other, both sublattices have almost the same contrast and the lattice looks hexagonal. Again A2 is to blame: it lines up with the three-fold lattice and enhances the Se sites on one side, but across the boundary the Mo and Se sites swap places, and the same A2 now evens out their contrast. “Evaluation of the contrast could lead one to deduce that there are two different materials on the opposite sides of the boundary, which in fact is not the case,” the authors write. Such artefacts do not need exceptionally strong aberrations: about 100 nm of A2 was the typical state after tuning the corrector of the FEI Titan at 80 kV in Ulm.
Correction needs special recordings
Ways to obtain truly aberration-free images exist. They all recover the object wave, which can then be handled much more freely than an intensity image. In off-axis electron holography, a biprism makes the object wave interfere with a reference wave, and the wave is recovered numerically from the interference pattern[12]; the Cs corrector improves its detection limit for small phase shifts considerably[13]. In exit wave reconstruction, a series of images at equally spaced foci is recorded and the object wave is worked out iteratively[14]. Once the wave is known, any residual aberration can be corrected[15]. All these methods, however, need special arrangements during recording and cannot be applied to a single image afterwards. “Such a situation can be quite frustrating if the microscopist has managed to capture an elusive target during imaging, but the resulting image is of poor quality,” the authors write. The box below compares the methods.
One image, one Fourier transform
The new method works on the image itself, not on the wave. This is possible when the object scatters weakly, so that the linear imaging theory holds[16]; 2D materials such as graphene approximately fulfil this condition at 80 kV[17]. The aberrations of the microscope are described by the wave aberration function χ(u), where u is the spatial frequency. Its terms fall into two groups: symmetric ones, which are the same at u and −u, such as defocus and Cs, and anti-symmetric ones, which change sign, such as A2 and B2. In the linear theory, the Fourier transform of the image holds every scattered beam twice, once directly and once as its mirror image from −u. The symmetric part of χ enters the two terms with opposite signs, the anti-symmetric part with the same sign. Multiplying the Fourier transform of the image by exp(iχas(u)) and transforming back therefore removes all anti-symmetric aberrations at once[18]:
The symmetric aberrations cannot be removed this way. Defocus and Cs make the contrast transfer vanish at certain frequencies, and information lost there cannot be recovered. A tilt of the sample is neglected, which is reasonable for flat 2D materials. To correct an image, the aberration coefficients and their directions must be known. The corrector measures them from a series of images at tilted beam[4]; other methods use focal or tilt series[19] or even a single lattice image[20]. Alternatively, the coefficients can be found by trial and error, by looking for the values that give, for example, the pristine graphene lattice its expected appearance. This must be done with care, because different combinations of amplitude and orientation of A2 can produce the same sublattice imbalance in MoSe2. The authors recommend the orientation in which the smallest amplitude removes the visible effect: if the real aberration points elsewhere, some of it remains, but the image is still improved and never over-corrected. In 2D materials with three-fold symmetry, A2 has a distinct effect on the lattice and can be told apart from B2. Microscope, samples and simulation are listed in the box below.
A test with known aberrations
To check the method, the team first used simulated images, where the aberrations are known exactly (Fig. 2). A MoSe2 flake with an embedded triangular mirror domain, simulated without anti-symmetric aberrations, looks the same inside and outside the domain, apart from the reversed order of darker and brighter sublattices. With A2 = B2 = 100 nm, both at 30°, the lattice looks completely different inside and outside, just as in the experiment. After the correction with the known values, the aberration-free image was back. Even with 10 µm of each, far beyond any real microscope, the scrambled image was restored, as the box below shows.
Hexagons regained
Then the method was applied to the experimental images of Fig. 1 (Fig. 3). For graphene, A2 = 150 nm at 17° was found by trial and error to remove the visible difference between the two sublattices: the lattice regained its hexagonal look, and the defect became easier to analyse. For MoSe2, 75 nm at 15° restored the expected appearance of the lattice on both sides of the mirror-twin boundaries. The uncorrected MoSe2 image had one advantage: the point defects were more visible in one of the domains, and the boundaries were easier to find. This resembles the known trick of moving away from Scherzer focus to make point defects stand out[23] (see Non-Invasive Electron Microscopy[24]). The anti-symmetric aberrations add no information, they only make features more evident to the eye, and since they only shift phases without changing the phase contrast transfer, they can be removed afterwards without any loss. The method can just as well strengthen them during processing where that helps.
One micrometre on purpose
In a second test, the team set strong A2 with the corrector on purpose, 1.0 ± 0.1 µm at 48° according to the corrector software, and imaged two places in the graphene sample (Fig. 4). In the first, a tilt grain boundary ran through the image, but its structure could not be analysed, and the lattice looked completely different on its two sides, because the two lattices were oriented differently to A2. In the second, a point defect could be seen, but its structure stayed hidden. Starting from the value of the corrector and fine-tuning by eye, the correction removed the effect of A2 in both images. The grain boundary then showed the 5–7 dislocations expected for a tilt grain boundary in graphene[25] (see Grain Boundary Migration[26]), and the point defect showed the pattern of a reconstructed divacancy[27]. How the values found by eye compare with the measurement of the corrector is shown in the box below.
Challenges named by the authors
The authors list the limits of the method. The sample must scatter weakly, which excludes a large number of commonly studied materials; 2D materials fulfil the condition only approximately. Symmetric aberrations such as defocus and Cs cannot be corrected, since the information lost at the zeros of the contrast transfer function is gone. The aberration coefficients must be known in advance, and the correction can only be as accurate as their measurement. Found by trial and error, they are not unique, because several combinations lead to the same image, and several strong residual aberrations at once make the search harder. Very large aberrations leave a margin at the image edges that cannot be trusted, and a tilt of the sample is neglected. Finally, the corrector still has to be reasonably well tuned during recording, otherwise the operator cannot recognise what is in the image: the divacancy in Fig. 4b was not recognisable before the correction.
Looser tuning, better images
“In fact, the anti-symmetric aberration coefficients can be freely adjusted post-situ using the presented method,” the authors conclude. Since the anti-symmetric aberrations can be removed afterwards, the corrector need not be perfectly tuned at all times. Looser requirements for these coefficients speed up the tuning, and because the time for tuning is limited by the drift of the corrector[7], a faster tuning can end in a better state of correction. The same procedure also estimates residual aberrations wherever it is known how an aberration-free image should look. The images were taken with the image-side aberration-corrected FEI Titan 80-300 at 80 kV at Ulm University.
Resource:
Lehtinen, O., Geiger, D., Lee, Z., Whitwick, M. B., Chen, M.-W., Kis, A., Kaiser, U. (2015).
Numerical correction of anti-symmetric aberrations in single HRTEM images of weakly scattering 2D-objects.
Ultramicroscopy 151, 130–135.
https://doi.org/10.1016/j.ultramic.2014.09.010
Update – June 12, 2015
The news article of April 1, 2015 ends above. Later work is listed below.
Coma moves the atoms
South China Agricultural University in Guangzhou and Zhejiang University in Hangzhou simulated HRTEM images of graphene at 80 kV with one non-round aberration at a time, each set to a phase shift of π/4 at 7.2 nm−1[28]. They cite the Ulm correction as the latest step in dealing with such aberrations. Their simulations confirm the Ulm reading of the graphene image: A2 and the three-lobe aberration D4 change only the brightness of the atoms, not their positions, so that the image resembles a sample with two kinds of atoms. All other low-order aberrations shift the apparent atom positions: two-fold astigmatism by up to 30 pm, and with coma the lattice still looks hexagonal, while every measured position is off. Atoms of the same sublattice shift by about the same distance. With this rule they corrected the bond lengths measured in an image taken on a JEOL JEM-ARM200CF with cold field emission gun at 80 kV.
Update – January 5, 2018
Correction at video rate
Osaka University built a system that reconstructs the electron wave in real time and now also corrects coma and three-fold astigmatism[29]. They cite the Ulm study, together with the earlier Ulm work on nanotubes[5], as evidence that residual coma and A2 distort the images of graphene and carbon nanotubes even in Cs-corrected microscopes. On an uncorrected Hitachi HF-2000 at 200 kV (Cs 0.8 mm), focal series of 128 images were reconstructed on a graphics processor. Coma and A2 were measured beforehand from the apparent two-fold astigmatism under tilted illumination, 1834 ± 230 nm and 308 ± 77 nm, and then removed from the wave; two-fold astigmatism was measured and corrected continuously, to ±0.4 nm and ±2°. All corrections, together with automatic focus tracking, ran at video rate, one frame every 1/30 s.
Update – June 16, 2020
The same trap in STEM
The AIST in Tsukuba, with partners in Xi’an, found the artefact of the Ulm MoSe2 boundaries again in scanning TEM[30]. On a JEOL 2100F with Delta correctors and a cold field emission gun at 60 kV, the two sides of an inversion domain boundary in monolayer MoS2 doped with Re looked different: one side showed the honeycomb of the 2H phase, the other a much weaker S sublattice, as if it were a different phase. Citing the Ulm study for residual aberrations, they traced this to the three-fold astigmatism of the probe together with a local tilt of the sample. In simulations, A2 of 50 nm already changed the S sublattice markedly with its direction, and in daily STEM work A2 lies at 50–100 nm. Rotating the sample by 60° in the microscope reversed which side looked wrong. They advise great caution when assigning phases of 2D materials from probe-corrected STEM images.
Update – September 9, 2020
False phases and a cure
KAUST in Thuwal, Saudi Arabia, with partners in Göttingen, showed the same effect on a Thermo Fisher Titan Themis Z at 80 kV in STEM[31]. They cite the Ulm study among the HRTEM work on residual aberrations and note that commercial probe correctors keep A2 below 100 nm, a value “considered acceptable”. In MoS2, A2 of 100–150 nm made parts of the 2H lattice look like the 1T phase; rotating A2 by 60° moved the false contrast to the other domain, and reducing it to 50 nm brought the honeycomb back on both sides. A monochromated beam with 60 meV energy spread suppressed the artefact further. Together with the Tsukuba study, this carries the Ulm warning from HRTEM over to STEM: residual three-fold astigmatism of the order of 100 nm, routine after tuning, is enough to fake a second phase in a 2D material.
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Haider, M., Rose, H., Uhlemann, S., Schwan, E., Kabius, B., Urban, K. (1998). A Spherical-Aberration-Corrected 200 kV Transmission Electron Microscope. Ultramicroscopy, 75, 53–60. https://doi.org/10.1016/S0304-3991(98)00048-5
-
Meyer, J. C., Kisielowski, C., Erni, R., Rossell, M. D., Crommie, M. F., Zettl, A. (2008). Direct Imaging of Lattice Atoms and Topological Defects in Graphene Membranes. Nano Lett., 8, 3582–3586. https://doi.org/10.1021/nl801386m
-
Krivanek, O. L., Chisholm, M. F., Nicolosi, V., et al. (2010). Atom-by-Atom Structural and Chemical Analysis by Annular Dark-Field Electron Microscopy. Nature, 464, 571–574. https://doi.org/10.1038/nature08879
-
Uhlemann, S., Haider, M. (1998). Residual Wave Aberrations in the First Spherical Aberration Corrected Transmission Electron Microscope. Ultramicroscopy, 72, 109–119. https://doi.org/10.1016/S0304-3991(97)00102-2
-
Biskupek, J., Hartel, P., Haider, M., Kaiser, U. (2012). Effects of Residual Aberrations Explored on Single-Walled Carbon Nanotubes. Ultramicroscopy, 116, 1–7. https://doi.org/10.1016/j.ultramic.2012.03.008
-
SALVE news (2012). Residual Aberrations in HRTEM. salve-project.de, March 18, 2012. https://salve-project.de/newspress/residual-aberrations-in-hrtem.html
-
Barthel, J., Thust, A. (2013). On the Optical Stability of High-Resolution Transmission Electron Microscopes. Ultramicroscopy, 134, 6–17. https://doi.org/10.1016/j.ultramic.2013.05.001
-
Krivanek, O. L., Stadelmann, P. A. (1995). Effect of Three-Fold Astigmatism on High Resolution Electron Micrographs. Ultramicroscopy, 60, 103–113. https://doi.org/10.1016/0304-3991(95)00050-B
-
Lehtinen, O., Vats, N., Algara-Siller, G., Knyrim, P., Kaiser, U. (2014). Implantation and Atomic-Scale Investigation of Self-Interstitials in Graphene. Nano Lett., 15, 235–241. https://doi.org/10.1021/nl503453u
-
SALVE news (2015). First Observation of Self-Interstitials in Graphene. salve-project.de, January 1, 2015. https://salve-project.de/newspress/first-observation-of-self-interstitials-in-graphene.html
-
SALVE news (2015). Se-Deficient 2D MoSe2. salve-project.de, March 11, 2015. https://salve-project.de/newspress/microstructure-and-properties-of-Se-deficient-2D-MoSe2.html
-
Lichte, H. (1986). Electron Holography Approaching Atomic Resolution. Ultramicroscopy, 20, 293–304. https://doi.org/10.1016/0304-3991(86)90193-2
-
Geiger, D., Lichte, H., Linck, M., Lehmann, M. (2008). Electron Holography With a Cs-Corrected Transmission Electron Microscope. Microsc. Microanal., 14, 68–81. https://doi.org/10.1017/S143192760808001X
-
Coene, W. M. J., Thust, A., Beeck, M. O. D., Dyck, D. V. (1996). Maximum-Likelihood Method for Focus-Variation Image Reconstruction in High Resolution Transmission Electron Microscopy. Ultramicroscopy, 64, 109–135. https://doi.org/10.1016/0304-3991(96)00010-1
-
Thust, A., Overwijk, M. H. F., Coene, W. M. J., Lentzen, M. (1996). Numerical Correction of Lens Aberrations in Phase-Retrieval HRTEM. Ultramicroscopy, 64, 249–264. https://doi.org/10.1016/0304-3991(96)00022-8
-
Buseck, P., Cowley, J., Eyring, L. (1989). High-Resolution Transmission Electron Microscopy and Associated Techniques. Oxford University Press, Oxford.
-
Lee, Z., Meyer, J. C., Rose, H., Kaiser, U. (2012). Optimum HRTEM Image Contrast at 20 kV and 80 kV – Exemplified by Graphene. Ultramicroscopy, 112, 39–46. https://doi.org/10.1016/j.ultramic.2011.10.009
-
Ishizuka, K. (2013). Phase Retrieval From Image Intensities: Why Does Exit Wave Restoration Using IWFR Work So Well?. Microscopy, 62, S109–S118. https://doi.org/10.1093/jmicro/dft005
-
Meyer, R. R., Kirkland, A. I., Saxton, W. O. (2002). A New Method for the Determination of the Wave Aberration Function for High Resolution TEM. 1. Measurement of the symmetric aberrations. Ultramicroscopy, 92, 89–109. https://doi.org/10.1016/S0304-3991(02)00071-2
-
Stenkamp, D. (1998). Detection and Quantitative Assessment of Image Aberrations From Single HRTEM Lattice Images. J. Microsc., 190, 194–203. https://doi.org/10.1046/j.1365-2818.1998.3000848.x
-
Algara-Siller, G., Lehtinen, O., Turchanin, A., Kaiser, U. (2014). Dry-Cleaning of Graphene. Appl. Phys. Lett., 104, 153115. https://doi.org/10.1063/1.4871997
-
Koch, C. T. (2002). Determination of Core Structure Periodicity and Point Defect Density Along Dislocations. PhD thesis, Arizona State University.
-
Lehtinen, O., Tsai, I. L., Jalil, R., Nair, R. R., Keinonen, J., Kaiser, U., Grigorieva, I. V. (2014). Non-Invasive Transmission Electron Microscopy of Vacancy Defects in Graphene Produced by Ion Irradiation. Nanoscale, 6, 6569. https://doi.org/10.1039/c4nr01918k
-
SALVE news (2014). Non-Invasive Electron Microscopy. salve-project.de, April 22, 2014. https://salve-project.de/newspress/non-invasive-electron-microscopy.html
-
Kurasch, S., Kotakoski, J., Lehtinen, O., Skákalová, V., Smet, J., Krill, C. E., Krasheninnikov, A. V., Kaiser, U. (2012). Atom-by-Atom Observation of Grain Boundary Migration in Graphene. Nano Lett., 12, 3168–3173. https://doi.org/10.1021/nl301141g
-
SALVE news (2012). Grain Boundary Migration. salve-project.de, May 10, 2012. https://salve-project.de/newspress/grain-boundary-migration-in-graphene.html
-
Banhart, F., Kotakoski, J., Krasheninnikov, A. V. (2010). Structural Defects in Graphene. ACS Nano, 5, 26–41. https://doi.org/10.1021/nn102598m
-
Lin, F., Jian, J., Ye, L., Jin, C. (2015). Effects of Non-Rotationally Symmetric Aberrations on the Quantitative Measurement of Lattice Positions in a Graphene Monolayer Using High-Resolution Transmission Electron Microscopy. Microscopy, 64, 311–318. https://doi.org/10.1093/jmicro/dfv025
-
Tamura, T., Kimura, Y., Takai, Y. (2018). Development of a Real-Time Wave Field Reconstruction TEM System (II): Correction of Coma Aberration and 3-Fold Astigmatism, and Real-Time Correction of 2-Fold Astigmatism. Microscopy, 67, 37–45. https://doi.org/10.1093/jmicro/dfx127
-
Hong, J., Koshino, M., Xu, H., Feng, Q., Suenaga, K. (2020). STEM Imaging Artifacts With Three-Fold Astigmatism in Monolayer Transition Metal Dichalcogenides. Appl. Phys. Lett., 116, 243102. https://doi.org/10.1063/5.0011782
-
Lopatin, S., Aljarb, A., Roddatis, V., Meyer, T., Wan, Y., Fu, J.-H., Hedhili, M., Han, Y., Li, L.-J., Tung, V. (2020). Aberration-Corrected STEM Imaging of 2D Materials: Artifacts and Practical Applications of Threefold Astigmatism. Sci. Adv., 6, eabb8431. https://doi.org/10.1126/sciadv.abb8431