High Resolution at 20 kV
July 1, 2010 – For the first time, an aberration-corrected transmission electron microscope delivers high-resolution images at only 20 kV. In the Carl Zeiss laboratory in Oberkochen, the first SALVE microscope, a Zeiss Libra 200 with a Cs corrector and a monochromator from CEOS, resolves the 314 and 272 pm lattice fringes of silicon and the lattice of single-layer graphene. The narrow energy spread of the monochromator made the difference. Microscopes at JEOL and AIST in Tsukuba (30 kV) and at Harvard (80 kV) confirm the same year that low voltage and atomic-plane resolution go together.
Slow electrons are kind to fragile samples, but they have always given blurred pictures. In July 2010, Ute Kaiser of Ulm University and Simone Pokrant of Carl Zeiss presented the first high-resolution images taken with an aberration-corrected transmission electron microscope at only 20 kV. Gerd Benner, head of TEM at Carl Zeiss in Oberkochen, and his colleagues recorded them in the Zeiss laboratory there, on a Zeiss Libra 200 fitted with a spherical-aberration corrector from CEOS in Heidelberg and a monochromator in the electron gun: the first microscope of the SALVE project. In single-layer graphene and in silicon, the images transfer lattice fringes down to 272 pm.
Why go down to 20 kV
An electron at 80 kV still carries enough energy to knock a carbon atom out of a nanotube or a graphene sheet. Below about 60 kV, such knock-on damage stops for graphite and nanotubes, and for most materials made of light atoms the threshold lies in the range 20–60 kV[1]. At lower voltage the atoms also scatter more strongly, so light elements give more contrast[2]. Ulm University had shown one year earlier, with a corrected FEI Titan, that carbon nanotubes which die within seconds at 300 kV survive for minutes at 80 kV, and that the next step needs still lower voltages[3] (see Microscopy at the Bottom[4]).
The price is sharpness. Round electron lenses suffer from spherical aberration Cs, which correctors have compensated since 1998[5], and from chromatic aberration Cc: electrons of slightly different energy are focused at different heights. This focus spread grows with the ratio of the energy spread to the beam energy, so at 20 kV and the same energy spread it is four times larger than at 80 kV[6]. A corrector that removes both Cs and Cc was the long-term aim of SALVE[1],[7]. In 2010, the team took the first step with what was available: a Cs corrector and a narrower energy spread.
A microscope set up for 20 kV
The instrument was based on the Zeiss Libra 200 platform, with a hexapole Cs corrector for the objective lens and an electrostatic monochromator for the electron gun, both from CEOS. For this study, the alignment of the whole column was optimized for high-resolution work at 20 kV. The box below lists microscope and samples and shows the instrument in the Zeiss laboratory in 2010 (Fig. 1).
Silicon and graphene at 20 kV
The two standard samples show what the corrected microscope can do at 20 kV (Fig. 2). In single-layer graphene, recorded with tilted illumination, a faint hexagonal lattice appears between patches of contamination. In silicon viewed along the [110] direction, the lattice planes run across the whole field of view. The 314 pm spacing of the {111} planes and the 272 pm spacing of the {200} planes are clearly resolved; an enlarged part of the same sample and its diffractogram show the spots of both (Fig. 3).
One component made the difference. “An essential component in achieving this result was the reduction of the energy spread of the source enabled by the integrated electrostatic monochromator,” the authors write. How much it matters at 20 kV follows from the focus spread, as the box below shows with the calculated limits.
One project, three partners
The development took place within Ulm University’s SALVE project (Sub-Ångström Low-Voltage Electron Microscopy), led by Ute Kaiser. Carl Zeiss Microscopy was the partner for the instrument, and CEOS in Heidelberg for the aberration corrector and the monochromator. The German Research Foundation (DFG) and the Ministry of Science, Research and the Arts (MWK) of Baden-Württemberg funded the work. The partners shared one goal, the authors write: “making it possible to study radiation-sensitive (soft and hard matter) materials at ultra-high-resolution by means of transmission electron microscopy (TEM).”
Checked against the expectation
The measured fringes match the known lattice and the coherence estimate, as the next box shows.
Other microscopes at low voltage
In the same months, two other groups pushed corrected microscopes to low voltage, in Tsukuba and at Harvard, and Berkeley had imaged graphene at 80 kV two years before. The boxes below compare their results with the 20 kV images from Oberkochen; the first one shows silicon [110] from Tsukuba and from Oberkochen side by side at the same scale (Fig. 4).
For graphene, the closest comparison is a single-layer membrane imaged at Berkeley in 2008, shown next to the 20 kV image at the same scale in the next box (Fig. 5).
Gerd Benner, who recorded the 20 kV images and supported the Ulm team at Zeiss, was also a co-author of the Harvard study of the same microscope type; the next box sets its 80 kV image next to the 20 kV one at the same scale (Fig. 6).
Challenges named by the authors
The note is short, but it names its conditions. The result depended on the monochromator; without a reduced energy spread, the corrected microscope would not have transferred the silicon fringes. The silicon sample had to be very thin, about 4 nm, and was prepared by focused ion beam. Only standard samples were shown: the actual goal, radiation-sensitive soft and hard matter at ultra-high resolution, was still ahead.
The way to sub-Ångström at low voltage
With 272 pm at 20 kV, the SALVE microscope was still far from the sub-Ångström resolution in the project’s name. But the images showed that a corrected column can work at 20 kV at all, and that the energy spread is the next limit. Lowering it further with a monochromator, or removing its effect with a Cc corrector, became the programme of the following years.
Resource:
Kaiser, U. A., Pokrant, S. (2010).
High Resolution at 20 kV – A Demonstration of Low-Voltage TEM Performance.
Carl Zeiss Microscopy.
https://salve-project.de/press/first_measurements_20_kv.pdf
Update – April 1, 2011
The news article of July 1, 2010 ends above. Later work is listed below.
The 20 kV prototype characterized
Ulm, CEOS, Carl Zeiss, Vienna and Nottingham published the full characterization of the 20 kV prototype[10] (see TEM at 20 kV[11]). The measured Cc was 1.2 ± 0.1 mm. At 20 kV, chromatic aberration limited the resolution to about 0.52 nm without the monochromator, at 0.75 eV, and to 0.25 nm with it, at 0.17 eV; Young’s fringes reached 440 pm and 240 pm, respectively. This confirms the estimate above. Silicon fringes of 272 pm and graphene fringes of 213 pm were resolved, and C60 fullerenes inside a nanotube withstood about a hundred times the dose that destroys them at 80 kV.
Update – April 29, 2014
Down to 15 kV in Tsukuba
A Cs-corrected STEM at 15 keV compares with a Cs-corrected TEM at 20 kV. The JEOL and AIST team went further down with its cold field emission gun and higher-order corrector[12]. At 15 kV, a TEM image of silicon [110] showed fringes of 192 pm, with spots in the power spectrum at 111 pm, and ADF STEM resolved the {111}, {002} and {220} planes of 314, 272 and 192 pm. Side by side at the same scale, the 15 kV STEM image and the 20 kV TEM image of the Zeiss note show the same silicon lattice (Fig. 7).
Update – August 9, 2016
Cc corrected from 20 to 80 kV
After Carl Zeiss had left the project in 2014[13], CEOS, FEI (today Thermo Fisher Scientific) and Ulm put into operation the corrector that SALVE had aimed for, which removes both Cs and Cc[14] (see SALVE Cc Correction Realized[15]). The SALVE III microscope resolves atoms at every voltage from 20 to 80 kV. Its information limit lies between 76 pm at 80 kV and 139 pm at 20 kV, with 90 pm at 40 kV. At 20 kV, Young’s fringes now reach far beyond the 240 pm of the SALVE I prototype (Fig. 8).
With Cc corrected, the comparison with scanning microscopes has changed. A C3/C5-corrected STEM at 60 keV compares with a Cc/Cs-corrected TEM at 80 kV. In 2010, the Nion UltraSTEM resolved the single boron and nitrogen atoms of a boron nitride monolayer at 60 kV, with a probe of about 120 pm[16]. At 80 kV, where SALVE III transfers information to 76 pm[14], its phase-contrast image of graphene shows the honeycomb sharply over a large area, every carbon atom bright[17] (Fig. 9a, b). Even at 30 kV, half the Nion voltage, a raw SALVE III image of MoS2 shows its lattice at least as clearly (Fig. 9c)[14]. At 15 kV, the JEOL Triple C showed the lattice fringes of h-BN only where the sample was three or more layers thick (Fig. 9d)[12].
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Rose, H. H. (2009). Future Trends in Aberration-Corrected Electron Microscopy. Phil. Trans. R. Soc. A, 367, 3809–3823. https://doi.org/10.1098/rsta.2009.0062
-
Bell, D. C., Russo, C. J., Benner, G. (2010). Sub-Ångstrom Low-Voltage Performance of a Monochromated, Aberration-Corrected Transmission Electron Microscope. Microsc. Microanal., 16, 386–392. https://doi.org/10.1017/S1431927610093670
-
Kaiser, U., Chuvilin, A., Meyer, J. C., Biskupek, J. (2009). Microscopy at the Bottom. In: MC2009 – Microscopy Conference, Graz, Vol. 3: Materials Science, 379–384. https://doi.org/10.3217/978-3-85125-062-6-379
-
SALVE news (2009). Microscopy at the Bottom. salve-project.de, August 30, 2009. https://salve-project.de/newspress/microscopy-at-the-bottom.html
-
Haider, M., Uhlemann, S., Schwan, E., Rose, H., Kabius, B., Urban, K. (1998). Electron Microscopy Image Enhanced. Nature, 392, 768–769. https://doi.org/10.1038/33823
-
Reimer, L., Kohl, H. (2008). Transmission Electron Microscopy: Physics of Image Formation, 5th ed. Springer, New York. https://doi.org/10.1007/978-0-387-40093-8
-
Haider, M., Müller, H., Uhlemann, S., Zach, J., Loebau, U., Hoeschen, R. (2008). Prerequisites for a Cc/Cs-Corrected Ultrahigh-Resolution TEM. Ultramicroscopy, 108, 167–178. https://doi.org/10.1016/j.ultramic.2007.07.007
-
Sasaki, T., Sawada, H., Hosokawa, F., Kohno, Y., Tomita, T., Kaneyama, T., Kondo, Y., Kimoto, K., Sato, Y., Suenaga, K. (2010). Performance of Low-Voltage STEM/TEM With Delta Corrector and Cold Field Emission Gun. J. Electron Microsc., 59, S7–S13. https://doi.org/10.1093/jmicro/dfq027
-
Meyer, J. C., Kisielowski, C., Erni, R., Rossell, M. D., Crommie, M. F., Zettl, A. (2008). Direct Imaging of Lattice Atoms and Topological Defects in Graphene Membranes. Nano Lett., 8, 3582–3586. https://doi.org/10.1021/nl801386m
-
Kaiser, U., Biskupek, J., Meyer, J. C., et al. (2011). Transmission Electron Microscopy at 20 kV for Imaging and Spectroscopy. Ultramicroscopy, 111, 1239–1246. https://doi.org/10.1016/j.ultramic.2011.03.012
-
SALVE news (2011). TEM at 20 kV. salve-project.de, April 1, 2011. https://salve-project.de/newspress/transmission-electron-microscopy-at-20-kv.html
-
Sasaki, T., Sawada, H., Hosokawa, F., Sato, Y., Suenaga, K. (2014). Aberration-Corrected STEM/TEM Imaging at 15 kV. Ultramicroscopy, 145, 50–55. https://doi.org/10.1016/j.ultramic.2014.04.006
-
SALVE news (2014). ZEISS Leaves SALVE. salve-project.de, February 6, 2014. https://salve-project.de/newspress/zeiss-leaves-salve-project.html
-
Linck, M., Hartel, P., Uhlemann, S., et al. (2016). Chromatic Aberration Correction for Atomic Resolution TEM Imaging From 20 to 80 kV. Phys. Rev. Lett., 117, 076101. https://doi.org/10.1103/PhysRevLett.117.076101
-
SALVE news (2016). SALVE Cc Correction Realized. salve-project.de, August 9, 2016. https://salve-project.de/newspress/SALVE-CC-correction-realized.html
-
Krivanek, O. L., Chisholm, M. F., Nicolosi, V., et al. (2010). Atom-by-Atom Structural and Chemical Analysis by Annular Dark-Field Electron Microscopy. Nature, 464, 571–574. https://doi.org/10.1038/nature08879
-
CEOS GmbH. SALVE: Cc/Cs-corrected phase-contrast TEM image of a carbon monolayer (graphene) at 80 kV. ceos-gmbh.de, accessed October 19, 2026. https://www.ceos-gmbh.de/de/produkte/temkorrektoren/salve