Differential Contrast in STEM
December 20, 2016 – In scanning transmission electron microscopy, lithium atoms, or a single atom inside a stack of layers, are still hard to see. In 1977, Harald Rose proposed subtracting the signals of constructive and destructive interference to obtain pure phase contrast; early in 2016, a team at Lawrence Berkeley National Laboratory showed it at 300 kV on the TEAM I microscope. Researchers at Ulm University have now calculated it for a Cc/Cs-corrected STEM at 20 to 80 kV: graphene appears with 12 to 26 times the bright-field contrast, and with fifth-order correction a single atomic layer can be brought into focus.
Graphene, polymers and biological molecules consist of light atoms that hardly scatter electrons. In the scanning transmission electron microscope (STEM), where a focused probe is scanned across the sample, they are therefore nearly invisible: the annular dark-field detector collects only the few electrons scattered to high angles, and the bright-field detector sees a weak signal on top of a large background. In 1977, Harald Rose proposed a way out[1]. If the electrons in which the scattered and the unscattered wave interfere constructively are recorded separately from those in which they interfere destructively, and the two signals are subtracted, the background and the nonlinear parts of the image cancel. What remains is enhanced, linear phase contrast. Zhongbo Lee, Ute Kaiser and Harald Rose at Ulm University have now calculated what this differential contrast can do in a STEM equipped with a corrector for both spherical and chromatic aberration (Cc/Cs), as built for the SALVE project, at 20, 50 and 80 kV.
Almost four decades passed before the idea could be built. Earlier in 2016, a team at Lawrence Berkeley National Laboratory (LBNL), with partners at the University of Oregon and Gatan and with Rose as coauthor, showed it in an experiment on the TEAM I microscope at 300 kV[2]. With the Ulm calculation, Rose — who designed the first practical aberration corrector as well as the TEAM and SALVE correctors and joined Ulm University in 2010 as senior professor in the SALVE project[3] — carries his concept from the TEAM project at high voltage to the low voltages of SALVE.
Why STEM needs such a mode was set out by the LBNL group[2]. Phase-contrast TEM is the method of choice for biological samples, because it images weak phase objects at doses of 10 e/Å2 or less; but its contrast depends on the defocus, which changes the information transfer in a nonlinear way, so the images need careful inspection or simulation before they can be interpreted. STEM images are simpler to read, since the geometry of the detector alone sets the contrast. The annular dark-field detector gives incoherent contrast roughly proportional to the projected mass thickness, but it uses only the electrons scattered to high angles and therefore needs much more dose per area than TEM, which hits light elements such as carbon hardest. Annular bright field was designed for light atoms such as lithium and oxygen, but it is hard to optimize, because the ratio of its inner to outer angle is fixed by the physical detector. STEM has therefore served mostly materials science and only rarely biology.
Graphene: up to 26 times the contrast
The Ulm team calculated images of single-layer graphene for a Cc/Cs-corrected STEM, once in conventional bright-field mode and once in the differential mode, both with the same corrector (Fig. 3a). The calculation leaves out the damping of the contrast by partial coherence, so it shows the best case. In bright field, the graphene lattice is barely visible at all three voltages. In the differential mode, the honeycomb stands out clearly: its contrast reaches 12 times the bright-field contrast at 20 kV, 18 times at 50 kV and 26 times at 80 kV. With an aberration corrector and a detector matched to a phase plate, weak phase objects such as graphene can thus be imaged in STEM with decent contrast.
The trick lies in designing both ends of the microscope together. The phase-contrast transfer function of the corrected objective lens and the geometry of the detector are chosen to match a phase plate in the illumination, so that the spatial frequencies transferred with positive contrast add up to the same total as those transferred with negative contrast. When the two detector signals are subtracted, the nonlinear information then cancels. How Rose framed the idea in 1977 and how it works in practice is shown in the box above, with his split detector (Fig. 1) and the phase plate used at LBNL (Fig. 2).
Focusing through single atomic layers
The calculation goes one step further. If the corrector also removes the fifth-order spherical aberration, the largest usable aperture, and with it the illumination angle, grows. The depth of field becomes so shallow that a single atomic layer can be brought into focus. The team calculated a focal series at 50 kV through 16 layers of silicon viewed along the ⟨110⟩ direction, 2.8 nm thick, in which one silicon atom in the 8th layer had been replaced by germanium (Fig. 3b). Although the layers are only 190 pm apart, the germanium atom stands out sharply at a defocus of −12 Å and is blurred at −20 and −4 Å. In this mode, the differential contrast is extremely sensitive to depth, and focusing through a sample atomic plane by atomic plane may come within reach.
The line profiles in Fig. 3b, drawn along a row of atom pairs that contains the germanium atom, show how sharp this is. At −12 Å the germanium atom gives a single deep dip, to about 0.73 of the background, while its silicon neighbors reach only about 0.87. At −20 Å the same spot shows only a weak, split dip, and at −4 Å the germanium atom has vanished. A focus change of 8 Å, about four layer spacings, thus decides whether the atom is seen.
Whether the principle holds in a real microscope was shown at LBNL at 300 kV. The comparison of both studies is set out in the box below, which also holds the images from LBNL (Fig. 4).
Outlook
Differential contrast in an aberration-corrected STEM can be a powerful method for weakly scattering objects, and with a further improved Cc/Cs corrector it becomes so sensitive to depth that focusing through atomic planes may come into reach, the authors conclude. Fast pixelated detectors, as used at LBNL, now record the full diffraction pattern at every probe position, so the matched detector can be built in software. The idea Rose proposed in 1977 thus links the TEAM project at high voltage with SALVE at low voltage. The depth question it raises runs through later SALVE work, from depth-sectioned images of graphene[7] to electron diffraction at 20 to 300 kV[8].
A linear phase signal has a practical advantage: the image can be read directly, without comparison to image simulations, and it is recorded in focus, unlike conventional high-resolution phase contrast, which needs defocus[5]. The LBNL group sees the main use in beam-sensitive materials and in heterostructures of hard and soft matter, where light and heavy atoms must be seen in one image[5]. Rose named the price in 1977: the integrated signal of a split detector resolves slightly differently along and across the scan, and unless the intensity at the start of each scan line is known, the image is distorted across the scan[1].
Resource:
Lee, Z., Kaiser, U., & Rose, H. (2016).
Calculation of phase contrast in Cc/Cs-corrected STEM.
EMC 2016 Proc., 733–734.
https://doi.org/10.1002/9783527808465.EMC2016.6235
Update – September 26, 2018
The news article of December 20, 2016 ends above. Later work that builds on it is listed below.
Two years later, Zhongbo Lee, Ute Kaiser and Harald Rose published the full calculation behind the abstract[9]. They call the mode annular differential phase contrast (ADPC), since it belongs to the family of differential phase contrast methods and uses rings, and studied it for 20 to 80 kV, the range of the SALVE instrument. The phase plate need not be made of material: the corrector can mimic it by tuning the defocus and the spherical aberrations. Two obstacles have to go first. Without Cc correction, a focal spread of 4 nm wipes out the ADPC transfer of a C5-corrected STEM completely, so that it then performs even worse than a C3-corrected one. And the image spread caused by Johnson noise, thermally induced currents in the lenses and the corrector, halves resolution and contrast of a Cc/C5-corrected STEM at 30 pm. Once both are overcome, the calculated ADPC resolution at 30 kV is about 25 pm with Cc/C5 correction and 57 pm with Cc/C3 correction.
The depth of field was read from a single image: in a graphene sheet tilted by 54.4°, neighboring rows of atoms lie 1 Å apart in height, so the number of rows in focus gives the depth of field directly. At 30 kV the Cc/C5-corrected STEM, with an illumination angle of 155 mrad, reaches 2.9 Å in ADPC and 4.1 Å in incoherent bright field; with Cc/C3 correction it is 17.5 and 19 Å. A multislice calculation then added dynamical scattering: graphite 20 nm thick, 62 layers 3.35 Å apart, with one carbon atom in layer 31 and one in layer 32 replaced by germanium (Fig. 5). Focused on either layer, ADPC shows the germanium atom of that layer with high contrast, while incoherent bright field and high-angle annular dark field show both atoms faintly at once, and for the same dose with a much lower signal-to-noise ratio. From 16 layers of silicon in 2016 to 20 nm of graphite, the authors conclude that optical sectioning in ADPC could reveal the inner structure of thick samples without cutting them into slices.
Update – May 25, 2020
In Marburg, Shamail Ahmed, Kerstin Volz and co-workers from Giessen, Karlsruhe and BASF used a fast pixelated detector to make lithium visible in a battery material[10]. They recorded four-dimensional STEM data of the cathode material LiNiO2 on a double aberration-corrected JEOL JEM-2200FS at 200 kV with a pnCCD camera and formed the detectors afterwards in software: an annular bright-field (ABF) image from 8 to 16 mrad, and an enhanced ABF image from which the medium-angle bright field, 0 to 8 mrad, is subtracted. The authors trace this idea of combining intensities from different angular ranges back to Rose. Lithium does not show at all in the high-angle annular dark-field image, while the enhanced ABF image of a grain about 58 nm thick shows lithium, nickel and oxygen. At the transition from the layered to the rock-salt structure, the images revealed nickel on tetrahedral sites and oxygen displaced toward the particle surface. Because both virtual detectors come from the same data set, no two physical detectors have to be calibrated against each other.
Update – March 7, 2022
Five years later, Zhongbo Li (formerly Lee), Johannes Biskupek, Ute Kaiser and Harald Rose at Ulm University took up the integrated signal of Rose 1977 for thick samples[11]. In integrated differential phase contrast (IDPC), the difference signals of opposite detector segments are integrated into a map of the projected potential. Calculated focal series at 80 kV through a polyoxometalate particle in a carbon nanotube show the limit of the usual quadrant detector: the deeper the focus, the more the transfer depends on direction, and stripes make the image uninterpretable (Fig. 6). More sectors, cut in software from a pixelated detector, remove them: up to a thickness of 10 nm, a Cc/C3-corrected STEM needs at least 18 sectors, while a STEM corrected for C3 only gets by with 8 up to 25 nm, but at much lower resolution. As a rule, if the transfer peak of one sector pair is w degrees wide, at least 360/w sectors are needed. Integrated center-of-mass imaging (ICOM) shows the same stripes in a small field of view of 1.5 nm, but not at 5 nm.
Like Rose, the paper names two routes to phase contrast, which, unlike conventional STEM detectors, uses all electrons in the bright-field cone and so needs less dose. One is a Fresnel phase plate with a matched ring detector, made by tuning defocus and aberrations or, as at LBNL, from material; the Ulm group calls it annular differential phase contrast (ADPC). The other subtracts opposite segments of a sector detector, the route of IDPC. A pixelated detector forms any geometry, even 40 or 80 segments that could hardly be built, and also center-of-mass (COM) imaging, a split detector that exists only in software: each pixel is weighted by its momentum, so a thin sample gives the gradient of its potential, and integrating or differentiating the COM image gives the potential (ICOM) or the charge density. With Cc/C3 correction, focusing below the optimum raises the contrast at the expense of resolution. For ADPC the paper recalls the calculation of 2018: a depth of field below 3 Å with Cc/C5 correction at 30 kV and 155 mrad, provided all aberrations that are not round are removed.
The key is the depth of field. With Cc/C3 correction the illumination angle at 80 kV can be 62 instead of 30 mrad, and the depth resolution improves about fourfold, from 9.2 to 2.2 nm. A proof-of-principle experiment on an uncorrected Thermo Fisher Talos at 200 kV turned the quadrant detector into a 12-sector detector by rotating the sample holder between exposures. With the probe focused 20 nm deep in an InGaN/GaN cross-section about 30 nm thick, the 12-sector image resolved the quantum wells, while the quadrant image was uninterpretable and the annular dark-field image did not show them; the glue line of the cross-section, too, appeared as a bright interface only in the 12-sector image. With manual rotation and a 2k × 2k camera, atomic resolution was not reached; a pixelated detector should do better. The authors also see a use for biological samples: the strong transfer of low spatial frequencies gives good contrast without a large defocus or a phase plate.
Download this news article as PDF – for reading on paper, taking notes or filing.
-
Rose, H. (1977). Nonstandard Imaging Methods in Electron Microscopy. Ultramicroscopy, 2, 251–267. https://doi.org/10.1016/S0304-3991(76)91538-2
-
Ophus, C., Ciston, J., Pierce, J., Harvey, T. R., Chess, J., McMorran, B. J., Czarnik, C., Rose, H. H., Ercius, P. (2016). Efficient Linear Phase Contrast in Scanning Transmission Electron Microscopy With Matched Illumination and Detector Interferometry. Nat. Commun., 7, 10719. https://doi.org/10.1038/ncomms10719
-
SALVE news (2010). Harald H. Rose joins Ulm University in 2010. salve-project.de, June 1, 2010. https://salve-project.de/newspress/senior-professor-harald-h-rose-at-ulm-university.html
-
Hammel, M., Rose, H. (1995). Optimum Rotationally Symmetric Detector Configurations for Phase-Contrast Imaging in Scanning Transmission Electron Microscopy. Ultramicroscopy, 58, 403–415. https://doi.org/10.1016/0304-3991(95)00007-N
-
Yang, H., Ercius, P., Nellist, P. D., Ophus, C. (2016). Enhanced Phase Contrast Transfer Using Ptychography Combined With a Pre-Specimen Phase Plate in a Scanning Transmission Electron Microscope. Ultramicroscopy, 171, 117–125. https://doi.org/10.1016/j.ultramic.2016.09.002
-
SALVE news (2016). SALVE Cc Correction Realized. salve-project.de, August 9, 2016. https://salve-project.de/newspress/SALVE-CC-correction-realized.html
-
SALVE news (2018). 3D Imaging of Graphene. salve-project.de, August 30, 2018. https://salve-project.de/newspress/3d-imaging-of-2d-materials.html
-
SALVE news (2021). 3D Electron Diffraction of MoS2. salve-project.de, April 15, 2021. https://salve-project.de/newspress/3d-electron-diffraction-mos2.html
-
Lee, Z., Kaiser, U., Rose, H. (2019). Prospects of Annular Differential Phase Contrast Applied for Optical Sectioning in STEM. Ultramicroscopy, 196, 58–66. https://doi.org/10.1016/j.ultramic.2018.09.012
-
Ahmed, S., Bianchini, M., Pokle, A., Munde, M. S., Hartmann, P., Brezesinski, T., Beyer, A., Janek, J., Volz, K. (2020). Visualization of Light Elements Using 4D STEM: The Layered-to-Rock Salt Phase Transition in LiNiO2 Cathode Material. Adv. Energy Mater., 10, 2001026. https://doi.org/10.1002/aenm.202001026
-
Li, Z., Biskupek, J., Kaiser, U., Rose, H. (2022). Integrated Differential Phase Contrast (IDPC)-STEM Utilizing a Multi-Sector Detector for Imaging Thick Samples. Microsc. Microanal., 28, 611–621. https://doi.org/10.1017/S1431927622000289